The 

IKA

 Vibrax VXR basic is suitable for mixing liquids in bottles, flasks, test tubes and 

bowls with a maximum supported weight of 2 kg (including the attachment).
It is designed for use in laboratories. The motion of the agitation table or of the cont-
ainers placed on it is approximately circular.

For usage in accordance with requirements, the machine must be standing on a stable, 
even surface that is as slip-free as passible. In addition, care must be taken that qbjects 
in the vicinty are a sufficient distance away from the agitator and that they will not 
come too close to the agitator while it is in operation.
Make certain that the unit is standing clear on the spring loaded feed and that it is not 
pushing against anything anywhwere.

• Use 

- Please unpack the device carefully
- In the case of any damage a detailed report must be set immediately (post, rail  
  or forwarder)

• Range 

of 

use

IKA

 Vibrax VXR basic

- Operating instructions

With the purchase of this device, you have acquired a high-quality product. The design 
of the unit and its special shape ensure ease of handing and problem-free work. 
The speed-controlled external rotor asynchronous motor allows for infinitie speed 
adjustment in the range from 0 to 2200 rpm. Electronic motor control holds the set 
speed constant even if the weight of the material on the surface increases. The heat 
given off by the motor can be used to heat up the holding surface for the agitation 
containers.

The motion of the agitation table with the mass to be agitated exerts a reactive 
force on the drive. This force is comparable to an imbalance. As a result, there is an 
imbalance mass on the motor that compensates for a part of the imbalance forces. 
Because of the different mass of each agitation material, there is a residual imbalance. 
This imbalance has the tendency to move the unit back and forth and to move it 
around on the adapter surface. This is largely prevented by the spring-loaded feet. 
The vibration energy is stored in the springs and directed back into the machine. If the 
spring system, the overall mass of the device and the speed that is set form a system 
capable of oscillating, the oscillating motion of the device may result in resonance. Do 
not operate the device at a “critical speed” such as this. Either reduce the speed and 
agitate at a lesser intensity or pass through the critical as possible if you would like to 
work at a higher speed or greater agitation intensity.

Check whether the voltage specified on the type plate matches the mains voltage 
available.
Place the selected attachment on the agitation table and tighten it securely with the 
three clamping screws. Make certain the agitation material is securely in place.
Before tuning on the machine, adjust the turn dial for controlling the spped so that it 
is against the left stop.
If these conditions are met, the device is ready to operate after plugging in the mains 
plug. If these procedures are not followed, safe operation cannot be guaranteed and/or 
the equipment may be damaged.
Please note the ambient conditions indicated in the Technical details (temperature, 
and relative humidity).
After the device has been turned on, the control light comes on three times.

  8

Unpacking

Proper use

Useful facts

Commissioning

Summary of Contents for Vibrax VXR basic

Page 1: ...IKA Vibrax VXR basic 20000005403 VXR_basic_122017 ...

Page 2: ...ontens evenly distributed and well secured Because of improper loading and the position of the center of gravity dynamic forces may arise during the agitation process that cause the shaker to move about on the table In this case reduce the speed immediately Additional risk may be present for the user if flammable materials are used during the agitation process or if mechanical agitation energy is ...

Page 3: ...active force on the drive This force is comparable to an imbalance As a result there is an imbalance mass on the motor that compensates for a part of the imbalance forces Because of the different mass of each agitation material there is a residual imbalance This imbalance has the tendency to move the unit back and forth and to move it around on the adapter surface This is largely prevented by the ...

Page 4: ... circuit and the control light flashes To eliminate the error the weight of the material must be reduced allow the device to cool off The device must be turned off and backon again The IKA Vibrax VXR basic is maintenace free For cleaning disconnect the main plug It is subject only to the natural wear and tear of components and their statistical failure rate If the three silicon hoses that protect ...

Page 5: ...on stroke mm 4 Ø Shaking motion horizontal circular Perm duration of operation 100 Perm ambient temperature C 5 to 40 Perm relative humidity 80 Prozection class acc DIN 40 050 IP 21 Protection class II Overvoltage category II Contamination level 2 Operation at a terrestrial altitude m max 2000 above sea level Max load kg 2 including attachment Dimensions W x D x H mm 157 x 247 x 130 without attach...