Page 14
WARNING
Electrical Connection
Connect the generator to an appropriate 100-230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician. If a substitute mains supply cord is used, ensure that
it has an adequate rating. Failure to do so could cause damage to the generator.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check that your electrical supply falls within the specified range, see
below.
EARTH/GROUND (E):-
Green & Yellow or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements are 100-230VAC n/- 10% depending on chosen model.
However, running continuously at voltages outside this is not recommended. Extended
periods at extremes can have a detrimental effect on the operation and life of the
generator.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
Unit is fully pneumatic and can function without electricity.
If the unit is not connected to a power supply the pressure status indicator feature
will not be available and the LED lights will not function. The unit will perform all
other functions.
WARNING
Summary of Contents for Solaris XE
Page 23: ...Page 23...