Installation
Generator Environment
The generator should be situated in a well ventilated environment and positioned such
that it can be easily disconnected if required. The generator is designed for indoor use
only. It should be installed adjacent to, underneath or in between the application(s) it
is supplying (maximum weight bearing up to 100kg) or it can be mounted onto a wall
horizontally; refer to installation guide for mounting details. If this is not convenient
then the unit can be sited elsewhere, however, consideration should be made of the
lengths of pipe runs as pressure drops can result from extended runs of pipe.
The performance of the generator (like all sophisticated equipment) is affected by
ambient conditions. Note should also be taken to the proximity of Air Conditioning
outlets. These can sometimes give rise to “pockets” of air with high relative humidity.
Operation of the unit within such a pocket could adversely affect its performance.
Consideration should also be given to the air flow around the unit. It is recommended
that an air gap of 75mm (3”) should be maintained down both sides and at the rear of
the unit. Please refer to the drawing on the next page for the general dimensions of the
unit.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)