Page 23
Normal Operation
Genius XE generators require minimal operator input. So long as the generator is
installed as described in earlier sections and is serviced in accordance with the
specified maintenance recommendations (see Service Requirements), then it will
operate in accordance with the demands of connected applications.
The generator will automatically produce the factory set pressure as detailed in the
technical Specifications. The outlet flow rate will vary to satisfy customer demand up
to a maximum flow rate in normal operating conditions of 35 LPM (Genius XE 35) or 70
LPM (Genius XE 70).
Genius XE generators are variable-purity systems, and will supply a higher purity of
nitrogen at lower flow rates, with a minimum purity of 95% nitrogen at the maximum
rated output flow.
Eco Mode
To reduce energy consumption and minimise component wear, Genius XE generators
feature a smart “Eco-Mode” which ensures gas is produced as is necessary to satisfy
demand.
When demand from the instrument stops the Genius XE generator will stop. If demand
from the application starts again, the system will detect the demand for gas and will
automatically restart. Additionally, if the flow demand from the generator is below its
maximum capacity, the generator will reduce the number of running compressors to
the minimum required.
N
2
GENIUS XE 35
S/N: XXXXXXXXXXXX
Nitrogen:
XXXX
psi
Home Screen
This is the main home screen in normal operation.
Idle Screen
The Idle Screen is displayed if no system changes occur
within a 20 Min period, is not running in Eco mode or
screen is not touched.
The screen is removed if the system changes or the
screen is touched and returns to the relevant home
screen.
Summary of Contents for Genius XE 70
Page 34: ...Page 34 Notes...
Page 35: ...Page 35 Notes...