Technical Specification
Environment
XE 35
C
XE 70
Minimum Operating Ambient Temperature
5°C / 41°F
Maximum Operating Ambient Temperature
35°C / 95°F
Maximum Altitude
3000m
Maximum Relative Humidity
80% @ 35°C
Minimum Storage Temperature*
-20°C / -4°F
Maximum Storage Temperature*
60°C / 140°F
Generator Outlets
Max Gas Outlet Pressure
116 psi / 8 bar
Max Gas Outlet Flow Rate
†
35 L/min
70 L/min
Purity
Up to 99.5%
Particles
<0.01
ɥ
m
Phthalates
Phthalate & BHT Free
Suspended Liquids
None
Hydrocarbon Removal
<1ppm NMHC
Gas Outlets
1 x 1/4" BSPP
Drain Outlet
1 x ¼" BSPP
Pressure Gauges/Displays
2
Start-Up Time
30 mins
Electrical Requirements
Voltage
120V±5% / 230V±10%
230V±10%
Frequency
120V 60Hz / 230V 50/60Hz
230V 50/60Hz
Current
12A @ 120V / 8A @ 230V
12A @ 230V
Input Connection
C20 Plug
Power Cord
C19 Socket
Circuit Breakers
230V 6A MCB, 2A MCB
120V 10A MCB, 2A MCB
Pollution Degree
2
Insulation Category
Class 1
Transient Over Voltages
Over Voltage Category II
General
Dimensions cm (inches) H x W x D
650 x 570 x 710 mm
(25.6 x 22.5 x 28'')
1000 x 570 x 710 mm
(39.4 x 22.4 x 28.0'')
Generator Weight Kg (lbs)
92 Kg (202.9 lbs)
147 Kg (324.1 lbs)
Shipping Weight Kg (lbs)
129 Kg (284.4 lbs)
182.5 Kg (402.4 lbs)
Noise Level
‡
55.5 dBA
59.3 dBA
* Note: Please ensure Generator is situated in a well ventilated environment.
†
Note: Flows in LPM are expressed as normalised volumes at 101.3 kPa, 20°C
‡
Note: Noise level expressed as SPL (Sound Pressure Level) measured at 1m from source in a reverberant chamber in accordance
with ISO 3741:2010.