Page 13
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere. Consideration should be made of the lengths of pipe runs as pressure drops
can result from extended runs of pipe. See page 21 for guidance on tubing lengths
greater than 3m.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the air
flow around the unit.
An air gap of 75mm (3”) should be maintained down both sides
and 100mm (4") at the rear of the unit.
Please refer to the drawing opposite for the
general dimensions of the unit.
Please ensure generator is situated in a well ventilated environment and is positioned to
permit easy diconnection if required.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)
Operation in High Ambient Temperatures
Genius XE generators are designed to supply the rated flow, purity, and pressure in
most laboratory environments, however high ambient temperatures can affect the
performance of the generator. Output flow capabilities may be restricted in high
ambient temperatures, or areas with insufficient air circulation.