background image

page 9 

Single-point adjustment with 50% humidity standard 

For  the  adjustment  the  appropriate  calibration  equipment  as  well  as 
calibration ampoules resp. humidity standards of 50 % r.h. are required.  

Proceedings 

Preparation 

To  ensure  as  good  as  possible  inspection  results  it  is  essential  that  the 
measuring  device,  the  calibration  equipment  and  the  calibration  ampoules 
have approximately the same temperature.  

This temperature has to be between 20°C and 26°C. 

The best way to ensure the same temperature of the different components is 
to  store  all  components  together  in  a  room  with  only  small  temperature 
fluctuations minimum over night – better for 24 hours.

 

Components of calibration equipment 

In  this  image  you  can  see  the  components  of  the 
calibration  equipment  and  a  calibration  ampoule 
with humidity standard.  

Assembly of calibration equipment 

1. Put in the first gasket ring in the upper part of

the calibration device.

2. Push in the sword sensor in the upper part as shown in the picture.
3. Now put the second gasket ring into the upper part.
4. Lay in the textile pad in the bottom part of the calibration device, and

pour the humidity standard carefully at the textile pad.

Summary of Contents for RH5

Page 1: ...User Manual RH5 Humidity Measuring Device with Sword Sensor Version 1 0_en PCE Instruments 2016...

Page 2: ...orm the sensor For heavy stacks and rolls please use the optionally available sword sensor holder and the tool for removing the sword sensor holder Measurement To switch on the instrument press the ke...

Page 3: ...ative humidity the excessive moisture would condense or form fog Dew point temperature The dew point indicates the temperature that the not completely saturated air has to reach in order to be complet...

Page 4: ...page 4 Design of the device batteries slot flipside measuring value calibration curve ON OFF button temperature selection of calibration curves air humidity and temperature sensor Option interface...

Page 5: ...r ON OFF Switch upper Switch lower Save Hold Autolog Watch saved data Enter suppliers data Menu Enter Switch upper Switch lower Exit Enter numbers Enter letters Next or right Left Yes No Shift OK Edit...

Page 6: ...ase press the button Supplier s data can also be entered on the PC subsequently Switching on the display lighting Press the key briefly the display lighting switches off automatically after approx 20...

Page 7: ...until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs or Auto Logs and confirm with again All saved logs will be transferred to...

Page 8: ...e cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please change the batteries and try again Press the button at your devi...

Page 9: ...temperature of the different components is to store all components together in a room with only small temperature fluctuations minimum over night better for 24 hours Components of calibration equipme...

Page 10: ...carefully and proceed as follows Conditioning the sensor To achieve best results let the sensor adjust for two hours The temperature has to be between 20 C and 26 C If the shown measuring value diffe...

Page 11: ...factory calibration as follows Reset to factory calibration 1 Press the button two times to reach the menu point Options 2 Select the menu item Reset using the button and confirm by pressing 3 Enter...

Page 12: ...ervals for checking the instrument depend on your operational demands and the required level of accuracy In general the drift of the sensor according to the degree of use constant humidity or use with...

Page 13: ...emove the batteries For eventual resulting damages we cannot provide any warranty In case of a faulty device please contact PCE Instruments Optional accessories PC interface for printing saved data on...

Page 14: ...page 14 Notes...

Page 15: ...h German French Italian Spanish Russian 4 pcs 1 5Volt AA Alkaline batteries for approx 1800 measurements after approx 4 minutes 30 mA with display lighting 128 x 64 matrix display with LED backlightin...

Page 16: ...ing To demonstrate the importance of temperature adjustment the table below shows measuring errors due to a temperature difference of only 1 C 1 8 F between the measuring instrument and the substance...

Reviews: