Positioning the instrument 

Insert the sword-sensor into the stack for only approx. 10 cm, and push it a 
few  more  centimetres  into  the  stack  every  10  seconds.  Let  your  RH5 
adequately  adjust  to  the  material  (at  least  5  minutes)  before  you  start  to 
measure, particularly when the material pile was stored at a different 
temperature than the device. When removing the sword-shaped sensor 
please ensure that there is no up or down movement, because this could 
deform the sensor. For heavy stacks and rolls please use the optionally 
available sword sensor holder and the tool for removing the sword sensor 
holder. 

Measurement 

To  switch  on  the  instrument,  press  the 

key  for 

three seconds. 
After showing the logo, the measuring window opens and 
the current temperature and moisture value is displayed.  

In the type selection menu the calibration curves can be 
changed  by  pressing 

  or 

.  The  calibration  curves 

saved in the device can be found in the following list.  

List of calibration curves 

calibration curve 

description 

unit 

measuring 

range 

rel. humidity 

relative humidity of air 

%RH 

0 to 100% 

dewpoint 

dewpoint 

°C 

or 

°F 

-55 to +60°C

resp. 

-67 to 140°F

Summary of Contents for RH5

Page 1: ...User Manual RH5 Humidity Measuring Device with Sword Sensor Version 1 0_en PCE Instruments 2016...

Page 2: ...orm the sensor For heavy stacks and rolls please use the optionally available sword sensor holder and the tool for removing the sword sensor holder Measurement To switch on the instrument press the ke...

Page 3: ...ative humidity the excessive moisture would condense or form fog Dew point temperature The dew point indicates the temperature that the not completely saturated air has to reach in order to be complet...

Page 4: ...page 4 Design of the device batteries slot flipside measuring value calibration curve ON OFF button temperature selection of calibration curves air humidity and temperature sensor Option interface...

Page 5: ...r ON OFF Switch upper Switch lower Save Hold Autolog Watch saved data Enter suppliers data Menu Enter Switch upper Switch lower Exit Enter numbers Enter letters Next or right Left Yes No Shift OK Edit...

Page 6: ...ase press the button Supplier s data can also be entered on the PC subsequently Switching on the display lighting Press the key briefly the display lighting switches off automatically after approx 20...

Page 7: ...until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs or Auto Logs and confirm with again All saved logs will be transferred to...

Page 8: ...e cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please change the batteries and try again Press the button at your devi...

Page 9: ...temperature of the different components is to store all components together in a room with only small temperature fluctuations minimum over night better for 24 hours Components of calibration equipme...

Page 10: ...carefully and proceed as follows Conditioning the sensor To achieve best results let the sensor adjust for two hours The temperature has to be between 20 C and 26 C If the shown measuring value diffe...

Page 11: ...factory calibration as follows Reset to factory calibration 1 Press the button two times to reach the menu point Options 2 Select the menu item Reset using the button and confirm by pressing 3 Enter...

Page 12: ...ervals for checking the instrument depend on your operational demands and the required level of accuracy In general the drift of the sensor according to the degree of use constant humidity or use with...

Page 13: ...emove the batteries For eventual resulting damages we cannot provide any warranty In case of a faulty device please contact PCE Instruments Optional accessories PC interface for printing saved data on...

Page 14: ...page 14 Notes...

Page 15: ...h German French Italian Spanish Russian 4 pcs 1 5Volt AA Alkaline batteries for approx 1800 measurements after approx 4 minutes 30 mA with display lighting 128 x 64 matrix display with LED backlightin...

Page 16: ...ing To demonstrate the importance of temperature adjustment the table below shows measuring errors due to a temperature difference of only 1 C 1 8 F between the measuring instrument and the substance...

Reviews: