background image

page 3 

Description of definitions 

Relative  air  humidity:   

indicates  the  relation  between  the  current  water 

vapour  pressure  and  the  maximal  possible  water  vapour  pressure  (called 
saturation vapour pressure) 
The  relative  humidity  shows  the  degree  the  air  is  saturated  with  water 
vapour. For example: 
50%  relative  humidity  indicates  that  at  the  current  temperature  and  the 
current  pressure  the  air  is  saturated  with  water  vapour  for  half  of  its  value, 
100%  relative  humidity  means  that  the  air  is  totally  saturated.  When  the  air 
has  more  than  100%  of  relative  humidity,  the  excessive  moisture  would 
condense or form fog. 

 

Dew point temperature:

  The dew point indicates the temperature that the 

not completely saturated air has to reach in order to be completely saturated 
with  water  vapour.  If  the  room  with  the  current  relative  humidity  is  cooled 
down to the dew point temperature, the water vapour begins to condense. 

Application range 

Within  the  normal  application  range  (normal  range)  the  accuracy  of  the 
device is as indicated. A long-term application beyond the normal application 
range  (max.  range),  particularly  at  an  air  humidity  of  more  than  80%,  can 
lead  to  higher  measuring  errors  (+3%  after  60  hours).  Back  in  the  normal 
application  range,  the  sensor  will  return  to  the  indicated  accuracy 
automatically.   

Summary of Contents for RH5

Page 1: ...User Manual RH5 Humidity Measuring Device with Sword Sensor Version 1 0_en PCE Instruments 2016...

Page 2: ...orm the sensor For heavy stacks and rolls please use the optionally available sword sensor holder and the tool for removing the sword sensor holder Measurement To switch on the instrument press the ke...

Page 3: ...ative humidity the excessive moisture would condense or form fog Dew point temperature The dew point indicates the temperature that the not completely saturated air has to reach in order to be complet...

Page 4: ...page 4 Design of the device batteries slot flipside measuring value calibration curve ON OFF button temperature selection of calibration curves air humidity and temperature sensor Option interface...

Page 5: ...r ON OFF Switch upper Switch lower Save Hold Autolog Watch saved data Enter suppliers data Menu Enter Switch upper Switch lower Exit Enter numbers Enter letters Next or right Left Yes No Shift OK Edit...

Page 6: ...ase press the button Supplier s data can also be entered on the PC subsequently Switching on the display lighting Press the key briefly the display lighting switches off automatically after approx 20...

Page 7: ...until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs or Auto Logs and confirm with again All saved logs will be transferred to...

Page 8: ...e cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please change the batteries and try again Press the button at your devi...

Page 9: ...temperature of the different components is to store all components together in a room with only small temperature fluctuations minimum over night better for 24 hours Components of calibration equipme...

Page 10: ...carefully and proceed as follows Conditioning the sensor To achieve best results let the sensor adjust for two hours The temperature has to be between 20 C and 26 C If the shown measuring value diffe...

Page 11: ...factory calibration as follows Reset to factory calibration 1 Press the button two times to reach the menu point Options 2 Select the menu item Reset using the button and confirm by pressing 3 Enter...

Page 12: ...ervals for checking the instrument depend on your operational demands and the required level of accuracy In general the drift of the sensor according to the degree of use constant humidity or use with...

Page 13: ...emove the batteries For eventual resulting damages we cannot provide any warranty In case of a faulty device please contact PCE Instruments Optional accessories PC interface for printing saved data on...

Page 14: ...page 14 Notes...

Page 15: ...h German French Italian Spanish Russian 4 pcs 1 5Volt AA Alkaline batteries for approx 1800 measurements after approx 4 minutes 30 mA with display lighting 128 x 64 matrix display with LED backlightin...

Page 16: ...ing To demonstrate the importance of temperature adjustment the table below shows measuring errors due to a temperature difference of only 1 C 1 8 F between the measuring instrument and the substance...

Reviews: