8.2.3 Linear calibration
Three different weights with large weight size differences should be used for the calibration. The scale
must hang freely whilst measuring.
To change the linear calibration, press
again.
“LoAd0“is displayed.
Press
to start the measurement.
Wait until the measurement is stable.
Press
again. “LoAd1” appears on the display
Press
repeatedly. The display shows 00000. With the arrow buttons, enter the weight which
you will hang on the scale first. If you press the F1 button then you can also set low values.
Confirm your entry by pressing
.
Wait until your measurement is stable.
Press
again. “LoAd2” appears on the display
If this value is sufficient, then switch the equipment off or press
to continue.
Press
repeatedly. 00000 appears on the display.
The second measuring unit can be entered with the arrow buttons. Also hang the second
measuring unit on the scale. With F1 small entries can also be entered.
Press
and wait for the value to become stable.
Press
. “LoAd3” is displayed.
If this entry is enough, then switch the equipment off or press
to carry on.
Press
repeatedly. The display shows 00000.
The third measuring unit can also be entered with the arrow buttons. Also hang the third
measuring unit on the scale. With F1 small entries can also be entered.
Press
and a further measurement will be carried out.
Wait for the reading to stabilize and then press
and the calibration has now been completed.