PCE Instruments PCE-CS 1000 N Manual Download Page 11

 

OPERATING INSTRUCTIONS 

 

11 

Custom settings 

Important: The following setting options should only be carried out by a suitably trained operator. Incorrect 
settings can make the scale unusable. The remote control is required to get to the custom settings

 

Press 

the 

“▲“button twice. “P0000“ appears on the display. The codes for the individual sections are to be 

entered here with help of the blue arrow button on the remote control. Confirm the code with 

8.1 

Equipment configuration 

 

 

The code for this configuration is “0258”. 

 

Confirm the code with 

. Scale is displayed. 

 

8.1.1  Change resolution 

 

Press 

again. 

“E (resolution)“ is displayed. 

 

The resolution can be changed with the arrow buttons on the remote control. 

 

To save the settings, press 

 

until “End” appears on the display. 

 

8.1.2  Automatic zero tolerance 

 

To set the automatic zero tolerance, press 

 

twice. “A (percent value)” is displayed. 

 

The tolerance can be changed with the arrow buttons on the remote control. Choose between 0 
(switched off), 2(±2% FS), 3(±3% FS), 4 (±4% FS), 10 (±10% FS), 20 (±20% FS) and 100 
(±100% FS)- 

 

To save the settings, press 

 

until “End” appears on the display. 

 

8.1.3  Manual zero tolerance 

 

Press 

 three times (percentage 

value)” is displayed. This function is not explained 

8.1.4  Zero coverage area tolerance 

 

Press 

 four times (percentage 

value)” is displayed. This function is not explained 

 

8.1.5  Not explained function 

 

Press 

 five times (percentage 

value)” is displayed. This function is not explained 

 

8.1.6  Zero setting 

 

To turn the Zero setting on and off, press 

 

six times. “5 (on/off)”. 

 

This function can be activated (on) or deactivated (off) 

with the arrow buttons “up” and “down”. 

 

To save the settings, press 

 

until “End” appears on the display. 

 

Summary of Contents for PCE-CS 1000 N

Page 1: ...US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire S...

Page 2: ...c switch off 10 6 2 br 3 Set background lighting 10 6 3 IdL Lighting duration 10 7 Troubleshooting 10 8 Custom settings 11 8 1 Equipment configuration 11 8 1 1 Change resolution 11 8 1 2 Automatic zer...

Page 3: ...the manufacturer and the operator adhere to certain requirements When using the measuring equipment no risk must be posed to the operator The operator or the person responsible for operating the measu...

Page 4: ...s to be allowed to be dropped or be pulled along the ground or pulled at an angle The maximum load as stipulated in the operating instructions MAX must never be exceeded this also includes the maximum...

Page 5: ...functioning correctly After every 12 months or after 50 000 measurements depending on which comes first Check all load bearing parts for example hairline cracks etc Only trained personnel are to carr...

Page 6: ...nge Tare range 100 F s Response time 1 second Safety load 150 On no account may this be exceeded Display LCD with background lighting Digit height 20 mm Ambient temperature 10 C 40 C Power supply 3 x...

Page 7: ...ontrol 1 x operating instructions 4 Equipment description 4 1 Measuring equipment 4 1 Power button 4 2 Infrared sensor 4 3 Display 4 4 Tare Zero button Display Shows the measured values Power button T...

Page 8: ...nts 4 8 Left 4 9 Setup Upwards Zero Move one value upwards Set the measuring value to zero Hold button Enter Saves the value Confirms the entry Downwards Moves one position to the right Unit change Ch...

Page 9: ...h the remote control 5 2 Zero Press the button until an acoustic signal sounds Z appears on the display The remote control can also be used to do this Press 0 Make sure the value is stable Otherwise E...

Page 10: ...ntil End appears on the display 6 3 IdL Lighting duration To set the lighting duration press three times ldl lighting duration appears on the display Press the T to set the lighting You can set it on...

Page 11: ...until End appears on the display 8 1 2 Automatic zero tolerance To set the automatic zero tolerance press twice A percent value is displayed The tolerance can be changed with the arrow buttons on the...

Page 12: ...ess until End appears on the display 8 1 9 Setting gravity To set the gravity press nine times value is displayed The value can be changed with the up and down buttons The lowest value is 0 000 and th...

Page 13: ...and for the 3 point calibration for example 1 3 2 3 or the full load The individual adjustment points can be freely set on the scale and can be adapted to the weights provided by the operator The cod...

Page 14: ...Press again LoAd2 appears on the display If this value is sufficient then switch the equipment off or press to continue Press repeatedly 00000 appears on the display The second measuring unit can be...

Page 15: ...h ElektroG R cknahme und Entsorgung von Elektro und Elektronikaltger ten our equipment can be returned We will either re use the equipment or it will be disposed of by a recycling company according to...

Reviews: