OPERATING INSTRUCTIONS
11
8
Custom settings
Important: The following setting options should only be carried out by a suitably trained operator. Incorrect
settings can make the scale unusable. The remote control is required to get to the custom settings
Press
the
“▲“button twice. “P0000“ appears on the display. The codes for the individual sections are to be
entered here with help of the blue arrow button on the remote control. Confirm the code with
.
8.1
Equipment configuration
The code for this configuration is “0258”.
Confirm the code with
. Scale is displayed.
8.1.1 Change resolution
Press
again.
“E (resolution)“ is displayed.
The resolution can be changed with the arrow buttons on the remote control.
To save the settings, press
until “End” appears on the display.
8.1.2 Automatic zero tolerance
To set the automatic zero tolerance, press
twice. “A (percent value)” is displayed.
The tolerance can be changed with the arrow buttons on the remote control. Choose between 0
(switched off), 2(±2% FS), 3(±3% FS), 4 (±4% FS), 10 (±10% FS), 20 (±20% FS) and 100
(±100% FS)-
To save the settings, press
until “End” appears on the display.
8.1.3 Manual zero tolerance
Press
three times (percentage
value)” is displayed. This function is not explained
8.1.4 Zero coverage area tolerance
Press
four times (percentage
value)” is displayed. This function is not explained
8.1.5 Not explained function
Press
five times (percentage
value)” is displayed. This function is not explained
8.1.6 Zero setting
To turn the Zero setting on and off, press
six times. “5 (on/off)”.
This function can be activated (on) or deactivated (off)
with the arrow buttons “up” and “down”.
To save the settings, press
until “End” appears on the display.