OPERATING INSTRUCTIONS 

 

13 

 

 

8.2 

Calibration 

If the scale is giving incorrect readings, it is possible to re-adjust them. The scale will permit the resetting 
or adjustment of the gravitational force. If the correct factor is entered, the scale will give an accurate 
weight reading. Should the correct gravitational force factor have already been entered and the scale is 
still giving incorrect readings, then the scale can be adjusted by up to three calibration points. Therefore 
there is also the option to carry out a 1 point / 2 point or even a 3 point adjustment. However if only one 
calibration weight is available, then the re-calibration can be concluded after the first calibration point. We 
suggest you to use ³/

 of the load for the 1 point calibration and for the 3 point calibration, for example 

1/3, 2/3 or the full load. The individual adjustment points can be freely set on the scale and can be 
adapted to the weights provided by the operator. 
 

 

The code for the configuration is “8416”  

 

Confirm the code by pressing

. “CAL” is displayed. 

 

Now press

“Un (KG/LB)” is displayed. 

 

Choose between KG (kg) and Lb (lb) w

ith the arrow buttons “up” and “down” 

8.2.1  Calibration gravity 

 

To change the calibration gravity press 

 

“G (value)” is displayed. 

 

The value can be changed with the arrow buttons on the remote control. Choose between 0.000 
and 9999. The default value is 9.794. The average value in Germany is 9.810. 

8.2.2  Max. Capacity 

 

To change the max. capacity, press 

 again. 5 numbers will be displayed. 

 

The value can be changed with the arrow buttons on the remote control. Choose between 0.0000 
and 99999. The default value is 01000. 

Do not exceed the maximum value of one ton!!!

 

 

To save the settings, press 

 

until “End” appears on the display. 

 

 

Summary of Contents for PCE-CS 1000 N

Page 1: ...US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire S...

Page 2: ...c switch off 10 6 2 br 3 Set background lighting 10 6 3 IdL Lighting duration 10 7 Troubleshooting 10 8 Custom settings 11 8 1 Equipment configuration 11 8 1 1 Change resolution 11 8 1 2 Automatic zer...

Page 3: ...the manufacturer and the operator adhere to certain requirements When using the measuring equipment no risk must be posed to the operator The operator or the person responsible for operating the measu...

Page 4: ...s to be allowed to be dropped or be pulled along the ground or pulled at an angle The maximum load as stipulated in the operating instructions MAX must never be exceeded this also includes the maximum...

Page 5: ...functioning correctly After every 12 months or after 50 000 measurements depending on which comes first Check all load bearing parts for example hairline cracks etc Only trained personnel are to carr...

Page 6: ...nge Tare range 100 F s Response time 1 second Safety load 150 On no account may this be exceeded Display LCD with background lighting Digit height 20 mm Ambient temperature 10 C 40 C Power supply 3 x...

Page 7: ...ontrol 1 x operating instructions 4 Equipment description 4 1 Measuring equipment 4 1 Power button 4 2 Infrared sensor 4 3 Display 4 4 Tare Zero button Display Shows the measured values Power button T...

Page 8: ...nts 4 8 Left 4 9 Setup Upwards Zero Move one value upwards Set the measuring value to zero Hold button Enter Saves the value Confirms the entry Downwards Moves one position to the right Unit change Ch...

Page 9: ...h the remote control 5 2 Zero Press the button until an acoustic signal sounds Z appears on the display The remote control can also be used to do this Press 0 Make sure the value is stable Otherwise E...

Page 10: ...ntil End appears on the display 6 3 IdL Lighting duration To set the lighting duration press three times ldl lighting duration appears on the display Press the T to set the lighting You can set it on...

Page 11: ...until End appears on the display 8 1 2 Automatic zero tolerance To set the automatic zero tolerance press twice A percent value is displayed The tolerance can be changed with the arrow buttons on the...

Page 12: ...ess until End appears on the display 8 1 9 Setting gravity To set the gravity press nine times value is displayed The value can be changed with the up and down buttons The lowest value is 0 000 and th...

Page 13: ...and for the 3 point calibration for example 1 3 2 3 or the full load The individual adjustment points can be freely set on the scale and can be adapted to the weights provided by the operator The cod...

Page 14: ...Press again LoAd2 appears on the display If this value is sufficient then switch the equipment off or press to continue Press repeatedly 00000 appears on the display The second measuring unit can be...

Page 15: ...h ElektroG R cknahme und Entsorgung von Elektro und Elektronikaltger ten our equipment can be returned We will either re use the equipment or it will be disposed of by a recycling company according to...

Reviews: