OPERATING INSTRUCTIONS
13
8.2
Calibration
If the scale is giving incorrect readings, it is possible to re-adjust them. The scale will permit the resetting
or adjustment of the gravitational force. If the correct factor is entered, the scale will give an accurate
weight reading. Should the correct gravitational force factor have already been entered and the scale is
still giving incorrect readings, then the scale can be adjusted by up to three calibration points. Therefore
there is also the option to carry out a 1 point / 2 point or even a 3 point adjustment. However if only one
calibration weight is available, then the re-calibration can be concluded after the first calibration point. We
suggest you to use ³/
4
of the load for the 1 point calibration and for the 3 point calibration, for example
1/3, 2/3 or the full load. The individual adjustment points can be freely set on the scale and can be
adapted to the weights provided by the operator.
The code for the configuration is “8416”
Confirm the code by pressing
. “CAL” is displayed.
Now press
.
“Un (KG/LB)” is displayed.
Choose between KG (kg) and Lb (lb) w
ith the arrow buttons “up” and “down”
8.2.1 Calibration gravity
To change the calibration gravity press
“G (value)” is displayed.
The value can be changed with the arrow buttons on the remote control. Choose between 0.000
and 9999. The default value is 9.794. The average value in Germany is 9.810.
8.2.2 Max. Capacity
To change the max. capacity, press
again. 5 numbers will be displayed.
The value can be changed with the arrow buttons on the remote control. Choose between 0.0000
and 99999. The default value is 01000.
Do not exceed the maximum value of one ton!!!
To save the settings, press
until “End” appears on the display.