443B101/443B102
21
Manual 19854 Rev. D
ECN 50523
Command: REF0
Purpose:
Turns off the internal sinusoidal reference oscillator.
Response:
The module responds with a “0” to tell the calling process the command has
been received.
Command String:
XYC0ZREF0
Description:
X
—
Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y
—
Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z
—
Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
REF0
—
Commands module to turn off the internal oscillator.
Response String:
“0”
Description:
This return indicates that the command was received by the card.
Example:
Command String:
“04C01REF0”
Response String:
“0”
The 443B101 in rack 0 and slot 4 is commanded to turn off the internal
oscillator.