443B101/443B102
33
Manual 19854 Rev. D
ECN 50523
Command: TEDD
Purpose:
Returns raw hex data stored in the TEDS device.
Response:
The module responds with the CRC followed by a 32 byte string of hex data.
Command String:
XYCOZTEDD
Description:
X
—
Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y
—
Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z
—
Module type for the target. Specifying "C01" in this argument addresses
a 443B101, spec
ifying “C02” addresses a 443B102.
TEDD
—
Commands the module to return raw hex data stored in the
TEDS device.
Response String:
CRC followed by 32 bytes of hex data.
Example:
Command String:
“04C01TEDD”
Return String:
C917D014D00E942200005C12EC64352D
8701000000000000000000000000000000
(where C9 is the CRC byte)
The 443B101 in rack 0 and slot 4 is commanded to turn return the raw hex data
stored in the TEDS device.