The creation of a table for automatic sessions is carried out starting from input parameters
set by the chosen measurement method
The automatic measurements expects the entry of the following parameters.
- Number of electrodes: defines the number of electrodes that will be used for the
measurement, the maximum configurable value is 32.
- Distance "a": defines the distance value "a" between two contiguous electrodes that must
be used by the operator to lay the electrodes.
- Number of levels: defines the maximum number of levels that will be applied to the
table construction algorithm. If the number of electrodes is too low, it may happen that
the maximum number of levels requested by the operator can not be reached, in this case
a value of "Real number of levels" will be shown on the right which will be lower than
the one set.
- Maximum factor "n": the parameter is used only in the Wenner-Schlumberger method,
and defines the maximum value that the "n" parameter can assume within the table's
construction algorithm.
Pressing the Ok button will end the automatic table creation procedure, the Gea RM1-PC
program will save a file in GPD proprietary format/owner format containing the
parameters used by the automatic measurements table construction algorithm in the hard
disk folder.
F
IGURE
44
-
CREATION PAGE OF A NEW AUTOMATIC PAGE
W
ENNER
-
S
CHLUMBERGER
Summary of Contents for GEA RM1 1.0
Page 4: ...4 TEST circuit diagram 67...
Page 5: ......
Page 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Page 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Page 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Page 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...