GEA RM1
29
Description of the systems interface in standalone mode
The commands are as follows.
-
> Execute: it allows the execution of the measure as described in paragraph 4.7,
with the difference that the result of the measure, being associated with a session,
can be accepted or not, and if so it will be memorized in the session. However, it
is also possible to perform a measurement that has already been carried out
previously, in this case accepting the result would overwrite the old value and
replace it with the new one.
-
>
Repeat: the command is active only in the case of measurements taken with the
SEV Schlumberger method, and for the last measurement of the session
performed correctly (therefore, if a measure did not reach the Sigma value
required, it would be possible to repeat the previous measure). It allows to add a
measurement to the session in order to have the same geometrical values for the
electrodes A and B with respect to the previous measurement, but to recalculate
the geometrical values for the electrodes M and N so as to bring them to the value
of a quarter with respect to the distance AB. Once added, this measurement will
be displayed and executed as a normal session measurement.
-
> The command for changing/modifying the geometric parameters does not have
a specific command line but is achieved by going to one of the following lines
which show the geometric values of the electrodes that you want to change and,
after pressing the "Set" key, proceed in entering a string (see paragraph
). The geometrical positions of the electrodes can be changed whether
they already have a predefined value (as part of a measurement session SEV
Wenner or SEV Schlumberger for measurement with defined electrode position),
and if they are valued at TBD (for example for a session of Generic SEV
measures). It is not possible to insert position values for the data evaluated at TBM
(in the case of SEV Schlumberger only for measurements with undefined
electrode position).
This command allows the operator, in addition to inserting the geometric values
of the actual position of the electrodes used for a generic measurement, also to
keep track of any deviations of the default position of the electrodes that are
necessary during the actual field measurement due to the presence of obstacles.
Viewing the parameters of an automatic session
The window allows you to view the parameters that were associated with an automatic
session during the session creation phase. The parameters come after the two possible
commands: "View Results" and "Run Session".
The command "See results" to visualize the results already acquired for this session, this
command is available only if the session has been completely or partially executed. By
Summary of Contents for GEA RM1 1.0
Page 4: ...4 TEST circuit diagram 67...
Page 5: ......
Page 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Page 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Page 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Page 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...