Below are the fields in which the operator must indicate the features of the table.
- Type of measure: the following cases can be chosen:
1. Manual: carries out the construction of a real table, with valorization of the individual
positions of the electrodes, for the execution of the session manually by the operator;
2. Automatic: foresees the storage on file of the parameter dei parametri caratteristici
della sessione di misura, so that during the automatic execution by the system and using
the multiplexer component it is possible to apply the chosen measurement algorithm;
3. Only PS: carries out the construction of a real simplified table to be able to execute the
Spontaneous Potential measurement from a system in manual mode.
- Measurement method: defines the laying method of the electrodes on the ground and
depends on the type of measurement previously chosen.
If the type of measure is Manual the choices of the method can be the following: SEV
Wenner, SEV Schlumberger, SEV Generic, Dipole-dipole Wenner Beta, Dipolo-dpolo
Wenner Gamma, Dipole-generic dipole and Polo-dipole.
If the type of measurement is Automatic, the method can be as follows: Wenner Alfa,
Wenner Beta, Wenner Gamma, Wenner-Schlumberger, Dipole-dipole.
If the type of measurement is the PS one, it will not be possible to choose a measurement
method.
The additional parameters change with the type of measurement method chosen, and will
be described for each indivudual case in the following paragraphs.
hereafter we will describe how to create tables for the different types, at the end of the
operation the Ok button must be pressed to save the table on the file.
Creation of a table for SEV manual measurements
The table of manual measurements implies that the operator must place the electrodes and
connect them to the Gea RM1 system for each single measurement, then he will proceed
in carrying out the single measurement by controlling it from the system.
By picking the Manual option as the Measurement type, it will be possible to choose only
among the methods available for this type.
Summary of Contents for GEA RM1 1.0
Page 4: ...4 TEST circuit diagram 67...
Page 5: ......
Page 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Page 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Page 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Page 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...