26
The following lines allow you to view some important information regarding the session.
List of all the menu items.
-
> See measurements: allows you to display the individual measurements of the
session if this is a manual or spontaneous Potential, or> See automatic parameters:
allows the display of the set parameters if the session is Automatic.
-
Type of measure: indicates if the session is related to Automatic, Manual or PS
measures.
-
Measurement method: indicates whether the session was constructed using
algorithms to choose the geometric position of the electrodes (SEV Wenner, SEV
Schlumberger, etc.).
-
Order: indicates the order in which the electrodes for the measurement must be
positioned (ABMN, AMNB, etc.).
-
Number of measures taken according to the number of measures planned by the
session.
-
Date of creation of the session.
-
Date of last edit of the session.
-
Full view / Simplified view: if the system is set to "Simplified view if possible"
(see paragraph
Simplified display of geometric measurement data
) the
command allows to choose for the specific session whether to display the
geometry information in integral format. If the system setting is "Integral View"
the command is not present.
By pressing the "See measurements" command, it is possible to view the individual
measurements contained within the session file.
Display of the measurements of the session
The number of the displayed measurement is shown in the screen title, also in the case of
complete measurements the value of R is shown (when there are no geometrical values
relating to the position of the electrodes for the measurement) or the value of Rho (Rho
= K * R, when the geometrical values associated with the position of the electrodes are
present for the measurement and it is possible to calculate the geometric coefficient K).
In this way it is possible to understand immediately whether the displayed measurement
has already been performed and with what result (R or Rho with numerical value), or if
the measurement has yet to be carried out (value of R or of Rho = "- ").
Summary of Contents for GEA RM1 1.0
Page 4: ...4 TEST circuit diagram 67...
Page 5: ......
Page 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Page 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Page 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Page 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...