Using the Water Quality Colorimeter
ater Quality Colorimeter
Use the colorimeter to perform a chemical analysis of water samples
using ezSample Snap Vial test kits.
Step 1: Pr
Step 1: Prepar
epare Samples for Analysis
e Samples for Analysis
exSample Snap Vial test kits can be used to test for concentrations of
Ammonia, Chlorine, Iron, Phosphate, and Nitrate. Refer to the
documentation for the ezSample Snap Vial Kit for instructions on how
to prepare the samples. When preparing the samples:
• Clean the ampoule using a lint-free wipe before inserting it into the
test chamber.
• Handle the ampoules by the tip to avoid getting fingerprints on the
ampoule body.
• Cover the ampoule in the test chamber with the Ampoule Cap to
keep out ambient light.
Figure 1. Ampoule cap covering an ezSample Snap Vial ampoule inserted
into the test chamber.
Step 2: Set Up Data Collection
Step 2: Set Up Data Collection
SP
SPARK
ARKvue
vue
1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open SPARKvue then click Sensor Data
Sensor Data
.
3. Select the Colorimeter that matches the device ID.
4. Deselect
the default measurement then select
a chemical
to measure.
Note:
Note: You may need to scroll down to see the additional
measurements.
5. Under Templates
Templates, select Digits
Digits.
P
PASCO Capst
ASCO Capstone
one
1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open Capstone then click Hardware Setup
Hardware Setup
.
3. Select the Colorimeter that matches the device ID.
Note:
Note: You may be prompted to calibrate the sensor. If you
want to calibrate now, click Calibrate
Calibrate and proceed to the
Calibrate the Sensor
Calibrate the Sensor instructions. Otherwise, click Cancel
Cancel.
4. Click Data Summary
Data Summary
.
5. Click and drag the chemical you are measuring to Digits
Digits
in the
Displays
Displays palette.
6. Click Data Summary
Data Summary
to close the panel.
Step 3: Calibr
Step 3: Calibrate the Sensor
ate the Sensor
SP
SPARK
ARKvue
vue
1. Click Hardware Setup
Hardware Setup
then click Configure Sensor
Configure Sensor
for the
Wireless Colorimeter.
Note:
Note: You may be prompted to calibrate the sensor. Click
Calibrate
Calibrate, if prompted.
2. Insert the calibration ampoule included with the ezSample Kit into
the Colorimeter and cover with the Ampoule Cap.
3. Click Calibrate
Calibrate. A green check
will indicate that the colorimeter
is calibrated.
4. Click Done
Done.
5. Close Hardware Setup
Hardware Setup.
P
PASCO Capst
ASCO Capstone
one
1. Click Calibration
Calibration
.
2. Select Wireless Colorimeter
Wireless Colorimeter from the list then click Next
Next.
3. Insert the calibration ampoule included with the ezSample Kit into
the Colorimeter and cover with the Ampoule Cap.
4. Click Calibrate
Calibrate. A green check
will indicate that the colorimeter
is calibrated.
5. Click Finish
Finish.
6. Click Calibration
Calibration
to close the panel.
Step 4: Collect Data
Step 4: Collect Data
SP
SPARK
ARKvue
vue
1. Insert an ampoule containing your a sample and cover with the
Ampoule Cap.
2. Click Start
Start
to begin data collection.
3. Click Stop
Stop
to end data collection.
4. Repeat the steps for additional test samples.
P
PASCO Capst
ASCO Capstone
one
1. Insert an ampoule containing a test sample and cover with the
Ampoule Cap.
2. Click Record
Record
to begin data collection.
3. Click Stop
Stop
to end data collection.
4. Repeat the steps for additional test samples.
Wireless Colorimeter and Turbidity Sensor | PS-3215
Product Guide