Using the Water Quality Colorimeter

ater Quality Colorimeter

Use the colorimeter to perform a chemical analysis of water samples
using ezSample Snap Vial test kits.

Step 1: Pr

Step 1: Prepar

epare Samples for Analysis

e Samples for Analysis

exSample Snap Vial test kits can be used to test for concentrations of
Ammonia, Chlorine, Iron, Phosphate, and Nitrate. Refer to the
documentation for the ezSample Snap Vial Kit for instructions on how
to prepare the samples. When preparing the samples:

• Clean the ampoule using a lint-free wipe before inserting it into the

test chamber.

• Handle the ampoules by the tip to avoid getting fingerprints on the

ampoule body.

• Cover the ampoule in the test chamber with the Ampoule Cap to

keep out ambient light.

Figure 1. Ampoule cap covering an ezSample Snap Vial ampoule inserted

into the test chamber.

Step 2: Set Up Data Collection

Step 2: Set Up Data Collection

SP

SPARK

ARKvue

vue

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open SPARKvue then click Sensor Data

Sensor Data

.

3. Select the Colorimeter that matches the device ID.
4. Deselect

the default measurement then select

a chemical

to measure.

Note:

Note: You may need to scroll down to see the additional

measurements.

5. Under Templates

Templates, select Digits

Digits.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open Capstone then click Hardware Setup

Hardware Setup

.

3. Select the Colorimeter that matches the device ID.

Note:

Note: You may be prompted to calibrate the sensor. If you

want to calibrate now, click Calibrate

Calibrate and proceed to the

Calibrate the Sensor

Calibrate the Sensor instructions. Otherwise, click Cancel

Cancel.

4. Click Data Summary

Data Summary

.

5. Click and drag the chemical you are measuring to Digits

Digits

in the

Displays

Displays palette.

6. Click Data Summary

Data Summary

to close the panel.

Step 3: Calibr

Step 3: Calibrate the Sensor

ate the Sensor

SP

SPARK

ARKvue

vue

1. Click Hardware Setup

Hardware Setup

then click Configure Sensor

Configure Sensor

for the

Wireless Colorimeter.

Note:

Note: You may be prompted to calibrate the sensor. Click

Calibrate

Calibrate, if prompted.

2. Insert the calibration ampoule included with the ezSample Kit into

the Colorimeter and cover with the Ampoule Cap.

3. Click Calibrate

Calibrate. A green check

will indicate that the colorimeter

is calibrated.

4. Click Done

Done.

5. Close Hardware Setup

Hardware Setup.

P

PASCO Capst

ASCO Capstone

one

1. Click Calibration

Calibration

.

2. Select Wireless Colorimeter

Wireless Colorimeter from the list then click Next

Next.

3. Insert the calibration ampoule included with the ezSample Kit into

the Colorimeter and cover with the Ampoule Cap.

4. Click Calibrate

Calibrate. A green check

will indicate that the colorimeter

is calibrated.

5. Click Finish

Finish.

6. Click Calibration

Calibration

to close the panel.

Step 4: Collect Data

Step 4: Collect Data

SP

SPARK

ARKvue

vue

1. Insert an ampoule containing your a sample and cover with the

Ampoule Cap.

2. Click Start

Start

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.

P

PASCO Capst

ASCO Capstone

one

1. Insert an ampoule containing a test sample and cover with the

Ampoule Cap.

2. Click Record

Record

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.