Using the Colorimeter

Using the Colorimeter

Use the colorimeter to measure the absorbance and transmittance of
six different wavelengths.

Step 1: Pr

Step 1: Prepar

epare Solutions for Analysis

e Solutions for Analysis

When preparing solutions for analysis:

• Use the lined cuvettes.
• Handle the cuvettes by the lined sides to avoid getting fingerprints

on the smooth sides.

• Fill cuvettes ¾ full with the test solution and cap the cuvette.
• Clean the smooth sides of the cuvette using a lint-free wipe before

inserting it into the test chamber.

Step 2: Pr

Step 2: Prepar

epare a Blank Calibr

e a Blank Calibration Cuv

ation Cuvette

ette

Fill one lined cuvette ¾ full with the solvent used to prepare the test
solutions. The solvent is used to calibrate the colorimeter.

Step 3: Set Up Data Collection

Step 3: Set Up Data Collection

SP

SPARK

ARKvue

vue

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open SPARKvue then click Sensor Data

Sensor Data

.

3. Select the Colorimeter that matches the device ID.
4. Select a Quick Start Experiment.

Absorbance and Transmittance

Absorbance and Transmittance
Measure absorbance and transmittance of a solution at six
different wavelengths on a bar meter display.

Beer's Law

Beer's Law
Plot absorbance versus concentration of known solutions to
determine the unknown concentration of a solution.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open Capstone then click Hardware Setup

Hardware Setup

.

3. Select the Colorimeter that matches the device ID.

Note:

Note: You may be prompted to calibrate the sensor. If you

want to calibrate now, click Calibrate

Calibrate and proceed to the

Calibrate the Sensor

Calibrate the Sensor section. Otherwise, click Cancel

Cancel.

4. Click Hardware Setup

Hardware Setup

to close the panel.

5. Select a Quick Start Experiment.

Absorbance and Transmittance

Absorbance and Transmittance
Measure absorbance and transmittance of a solution at six
different wavelengths on a bar meter display.

Beer's Law

Beer's Law
Plot absorbance versus concentration of known solutions to
determine the unknown concentration of a solution.

Step 4: Calibr

Step 4: Calibrate the Sensor

ate the Sensor

SP

SPARK

ARKvue

vue

1. Click Hardware Setup

Hardware Setup

then click

for the Wireless

Wireless

Colorimeter

Colorimeter.

Note:

Note: You may be prompted to calibrate the sensor. Click

Calibrate

Calibrate, if prompted.

2. Insert the blank calibration cuvette into the Colorimeter with the

smooth side facing the sensor and close the cover.

3. Click Calibrate

Calibrate. A green check

will indicate that the colorimeter

is calibrated.

4. Click Done

Done.

5. Close Hardware Setup.

P

PASCO Capst

ASCO Capstone

one

1. Click Calibration

Calibration

.

2. Select Wireless Colorimeter

Wireless Colorimeter from the list then click Next

Next.

3. Insert the blank calibration cuvette into the Colorimeter with the

smooth side facing the sensor and close the cover.

4. Click Calibrate

Calibrate. A green check

will indicate that the colorimeter

is calibrated.

5. Click Finish

Finish.

6. Click Calibration

Calibration

to close the panel.

Step 5: Collect Data (Absorbance and T

Step 5: Collect Data (Absorbance and Trransmittance)

ansmittance)

SP

SPARK

ARKvue

vue

1. Insert a cuvette containing your test sample with the smooth side

facing the sensor and close the cover.

2. Click Start

Start

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.

P

PASCO Capst

ASCO Capstone

one

1. Insert a cuvette containing your test sample with the smooth side

facing the sensor and close the cover.

2. Click Record

Record

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.

Step 5: Collect Data (Beer

Step 5: Collect Data (Beer''s Law)

s Law)

SP

SPARK

ARKvue

vue

1. Click Start

Start

to begin data collection.

2. Insert a cuvette containing a solution of known concentration

with the smooth side facing the sensor and close the cover.

3. Enter the known concentration in the table.
4. Click Keep

Keep

to record the Absorbance measurement.

5. Repeat steps 2-4 for each known solution.
6. Click Stop

Stop

to end data collection.

P

PASCO Capst

ASCO Capstone

one

1. Click Preview

Preview

to begin data collection.

2. Insert a cuvette containing a solution of known concentration

with the smooth side facing the sensor and close the cover.

3. Enter the known concentration in the table.
4. Click Keep Sample

Keep Sample

to record the Absorbance measurement.

5. Repeat steps 2-4 for each known solution.
6. Click Stop

Stop

to end data collection.

Wireless Colorimeter and Turbidity Sensor | PS-3215

Product Guide

3

012-15830B