background image

Using the T

Using the Turbidity Sensor

urbidity Sensor

Use the Turbidity sensor to measure the turbidity level (cloudiness) of
a water sample.

Step 1: Pr

Step 1: Prepar

epare Solutions for Analysis

e Solutions for Analysis

When preparing solutions for analysis:
• Use the smooth cuvettes that do not have lines.
• Fill cuvettes ¾ full with the test solution and cap the cuvette.
• Clean the smooth sides of the cuvette using a lint-free wipe before

inserting it into the test chamber.

Step 2: Pr

Step 2: Prepar

epare a Blank Calibr

e a Blank Calibration Cuv

ation Cuvette

ette

Deionized water is used for the first calibration point. Fill one smooth
cuvette ¾ full with deionized water and cap the cuvette.

Step 3: Pr

Step 3: Prepar

epare a 100 N

e a 100 NTU Calibr

TU Calibration Cuv

ation Cuvette

ette

The 100 NTU solution is used for the second calibration point.

Figure 2. A cuvette with the 100 NTU label placed

over the cap with the expiration date label attached.

1. Fill one smooth cuvette ¾ full with the 100 NTU solution and cap

the cuvette.

2. Place a 100 NTU cuvette label over the cuvette cap as shown in

the figure.

3. Place an expiration date label on the 100 NTU Cuvette label.

Note:

Note: The 100 NTU standard sample should provide

accurate results (±7%) up to the expiration date.

Step 4: Set Up Data Collection

Step 4: Set Up Data Collection

SP

SPARK

ARKvue

vue

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open SPARKvue then click Sensor Data

Sensor Data

.

3. Select the Colorimeter that matches the device ID.
4. Toggle the Wireless Turbidity Sensor

Wireless Turbidity Sensor switch from disable

to

enable

.

Note:

Note: You may need to scroll down the list of

measurements to see the sensor.

5. Under Templates

Templates, select Digits

Digits.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the Wireless Colorimeter and Turbidity sensor.
2. Open Capstone then click Hardware Setup

Hardware Setup

.

3. Select the Colorimeter that matches the device ID.

Note:

Note: You may be prompted to calibrate the sensor. If you

want to calibrate now, click Calibrate

Calibrate and proceed to the

Calibrate the Sensor

Calibrate the Sensor instructions. Otherwise, click Cancel

Cancel.

4. Toggle the Wireless Turbidity Sensor

Wireless Turbidity Sensor switch from disable

to

enable

.

5. Click Data Summary

Data Summary

.

6. Click and drag Turbidity (NTU)

Turbidity (NTU) to Digits

Digits

in the Displays palette.

7. Click Data Summary

Data Summary

to close the panel.

Step 5: Calibr

Step 5: Calibrate the Sensor

ate the Sensor

SP

SPARK

ARKvue

vue

1. Click Hardware Setup

Hardware Setup

then click

for the Wireless Turbidity

Wireless Turbidity

Sensor

Sensor.

Note:

Note: You may be prompted to calibrate the sensor. If so,

click Calibrate

Calibrate.

2. Insert the blank calibration cuvette into the test chamber and

close the cover.

3. Click Calibrate

Calibrate in item 3. A green check

will indicate that

calibration is complete.

4. Insert the 100 NTU Standard cuvette in the test chamber with the

arrow pointing toward the Turbidity sensor and close the cover.

5. Click Calibrate

Calibrate in item 4. A green check

will indicate that

calibration is complete.

6. Click Done

Done.

7. Close Hardware Setup

Hardware Setup.

P

PASCO Capst

ASCO Capstone

one

1. Click Calibration

Calibration

.

2. Select Wireless Turbidity Sensor

Wireless Turbidity Sensor from the list then click Next

Next.

3. Insert the blank calibration cuvette into the test chamber and

close the cover.

4. Click Calibrate

Calibrate. A green check

will indicate that the turbidity

sensor is calibrated.

5. Insert the 100 NTU Standard cuvette in the test chamber with the

arrow pointing toward the Turbidity sensor and close the cover.

6. Click Calibrate

Calibrate. A green check

will indicate that the turbidity

sensor is calibrated.

7. Click Finish

Finish.

8. Click Calibration

Calibration to close the panel.

Step 6: Collect Data

Step 6: Collect Data

SP

SPARK

ARKvue

vue

1. Insert a cuvette containing a test sample in the test chamber and

close the cover.

2. Click Start

Start

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.

P

PASCO Capst

ASCO Capstone

one

1. Insert a cuvette containing a test sample in the test chamber and

close the cover.

2. Click Record

Record

to begin data collection.

3. Click Stop

Stop

to end data collection.

4. Repeat the steps for additional test samples.

Wireless Colorimeter and Turbidity Sensor | PS-3215

Product Guide

5

012-15830B

Reviews: