ASLD2200 - Instruction Manual
13 Technical Specifications
13.1 General
Supply Voltage
115/230 VAC, 50/60 Hz, +10/-15%
Power Consumption
25 VA
Supply Fuse
Fuse 2.5A T 250V Antisurge 5mm x 20mm Glass
(Fitted to PCBA)
Operating Temperature
0 to 50°C
Storage Temperature
-20 to 70°C
Display
3 Digit LED
Sensing Speed
1 metre per minute
Manual Operation Speed
5 metres per minute
Accuracy
Dependent on depth of tank typ/- 5%*
Resolution
+/- 100 mm*
Terminals:
Max conductor cross section 2.5 mm
2
13.2 Alarm Output
Number of Output
1 Contact Closure
Alarm Types
Zero and/or 100% fixed
Relay Rating
0.5 A @ 100 VDC
13.3 Analogue Output
Type:
4-20 mA
Resolution:
Better than 1%
4-20mA Maximum Output Load: 500 Ohms
13.4 EMC
EMC:
EN 50081-1 (1992), EN 50082-1 (1994)
13.5 Physical Features
Dimensions:
350 x 490 x 382 mm (w
h
d)
Weight:
15 Kg
Environmental Rating:
IP55 Outer Enclosure
IP65 Electronics Housing
Material:
GRP
Cable Entry:
3
PG11
Mounting Legs
10 mm holes
Page 26 of 30
103370IM Issue 10 Date 19/03/2018