background image

 

 

5

 

10545100-Australia Rev. 

– 08/2018 

INTENDED USE: FLOWMETER FOR ANALGESIA DELIVERY SYSTEM 

The DIGITAL MDM mixer and accessories are analgesia gas mixing and delivery devices for use in medical procedures to 
relieve pain and anxiety. The devices permit the trained operator to adjust the percent of oxygen through a range of 100% 
to 30% (and thereby control the ratio of oxygen to nitrous oxide). The operator can also control the total volume of mixed 
gas up to 9.9 liters per minute. 

The Digital MDM Mixer may be used by a properly trained attending physician, dentist, or licensed professional provided 
they monitor the patient and themselves for any evidence of contraindications as stated in section titled Side effects and 
Contraindications on page 11. 

Nitrous Oxide should not be used in the following circumstances, this includes but is not limited to: 

 

Treatments that require a level of sedation greater than relative analgesia i.e. anaesthesia 

 

Patients who have maxillofacial injuries 

 

Patients who have a B12 deficiency 

 

Patients who have intracranial or intraocular pressure 

 

Treatments that require more than 24 hours of application. 

Continuous Flow Systems 

The Digital MDM Mixer continuous flow nitrous oxide and oxygen sedation system offers built in fail-safe mechanisms to 
prevent the flow of nitrous oxide without oxygen. All flowmeters feature easy adjustment of both flow rates and percentage 
N

2

O mixture (0-70%) allowing for predictable results across a wide variety of applications. 

The Digital MDM Mixer delivers a continuous flow of gas and must be used with scavenging breathing circuits and/or 
scavenging interface devices. 

Note: When used in any application gas may disperse to room air creating a potential exposure risk to Nitrous Oxide. A 
risk assessment should be conducted by facilities to assess exposure hazards and appropriate mitigation measures; eg: 
Adequate ventilation and N

2

O monitoring. (See: Room Ventilation on page 4)

 

Breathing Circuits 

Breathing circuits that can be used with any Porter and Matrx device include: 

 

Porter “double mask” autoclavable nasal hood breathing circuit 

 

Matrx by Parker Scavenging Nasal Mask breathing circuit with disposable or autoclavable nasal hoods 

 

Porter Silhouette disposable nasal hood and breathing circuit 

 

Porter Disposable medical breathing circuit with mask 

Various breathing circuits that may be used with Porter and Matrx devices must meet several criteria: 

 

22mm connection to the bag tee 

 

One-way valve (either in the bag tee or breathing circuit) to prevent rebreathing of exhalations 

 

Connection from the bag tee to a mask or nasal hood that the patient can comfortably wear or hold in place and 
achieve a proper seal 

 

Tubing or hose from the breathing circuit that can connect to a scavenging interface (19mm) and allow for adequate 
vacuum flow to wall evacuation system 

Gas Supply Connection 

After installation of the flowmeter, connect the nitrous oxide and  oxide supply  lines to the  Sleeve  Indexed System (SIS) 
fittings located on the back of the flowmeter unit. It is important that the regulators for both gases be set to give pressures 
in the range of 365.4 kPa ± 13.7 kPa. Confirm the absence of leaks at pressure connections on the unit (hose ends). A leak 
test using a soap/water solution can be used to determine if there is a leak. If there is a leak, soap bubbles will appear at 
the leaking location. Refer to FM-

916 Mobile “E” cylinder Stand Installation and Instructions for connecting gas supply to 

the flow meter and cylinder installation to mobile stand.

 

User Responsibility 

This product should not be repaired other than in accordance with written instructions provided by Matrx by Parker. The 
user  of  this  product  shall  have  the  sole  responsibility  for  any  malfunction  which  results  from  improper  use,  faulty 
maintenance, improper repair, damage, or alteration by anyone other than a Matrx by Parker authorized repairer Air Liquide 
Healthcare

. It is the user’s responsibility to ensure that the correct gas, at the correct pressure, is supplied to the fittings at 

the rear of the unit. 

 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Reviews: