PREPARATION FOR USE

 

CAUTION: Before using the Digital MDM Mixer, check the operation of the unit by performing the following tests. 

 

Performance  Checks:

  These  preliminary  tests  must  be 

conducted to ensure that your analgesia mixer is performing 
correctly.  If  the  unit  fails  any  one  of  the  following 
performance checks, the unit must be returned for service.

 

Machine Turn-On Test:

 Press the ON/OFF key, as shown 

in  Figure  1.  After  the  device  goes  through  an  initialization 
routine,  the  %  OXYGEN  and  LPM  FLOW  displays  should 
indicate 100% and 5 LPM. 

Alarm Indicator / Lamp Test:

 Press the alarm test / silence 

key,  as  shown  in  Figure  1,  and  observe  that  all  the  alarm 
readouts  illuminate.  The  %  OXYGEN  and  LPM  FLOW 
displays should show 8.8.8. and 8.8 respectively. The alarm 
readouts  should  illuminate  with  the  numerical  displays.  10 
bars  on  each  gas  should  illuminate.  The  audible  alarm 
should also sound. 

Oxygen  Failure  Test:

  Set  the  LPM  FLOW  display  to  9.9 

LPM.  Set  the  %  OXYGEN  display  to  50%.  Turn  off  the 
oxygen gas supply to the unit. When the gas runs out, check 
for each of the following: 

 

The OXYGEN FAILURE alarm illuminates 

 

The audible alarm sounds 

 

The flow indicators are off 

 

The LPM FLOW and % OXYGEN displays are off 

Turn on the oxygen gas supply, and the unit should return 
to normal operation. (The alarm may continue to chime for 
up to 20 seconds after restoration of gas pressure.) 

 

Nitrous  Failure  Test:

  Set  the  LPM  FLOW  display  to  9.9 

LPM.  Set  the  %  OXYGEN  display  to  30%.  Turn  off  the 
nitrous oxide gas supply to the unit. When the gas runs out, 
make sure each of the following occurs: 

 

The NITROUS FAILURE alarm illuminates 

 

The audible alarm sounds 

 

The  LPM  FLOW  display  indicates  the  oxygen  flow 
being delivered 

 

The % OXYGEN display is OFF 

 

The N

2

O gas flow indicator is OFF 

 

The O

2

 gas flow indicator shows at least 3 bars 

Turn on the nitrous oxide gas supply. The unit should return 
to normal operation. (The alarm may continue to chime for 
up to 20 seconds after restoration of gas pressure.) 

Indicated Flow Delivery Test:

 Set the % OXYGEN display 

to  100%.  Press  the  [

▼]

 

flow  key  and  hold,  as  shown  in 

Figure  1.  The  flow  rate  should  decrease  to  1.0  LPM.  An 
audible alarm sounds at the lower limit while the key is held. 
Press  the  [

▲]

 

flow  key  and  hold.  The  flow  rate  should 

increase  to  9.9  LPM.  An  audible  alarm  will  sound  at  the 
upper limit as long as the key is held. 

Indicated  Percentage  (%)  Delivery  Test:

  Set  the  LPM 

FLOW  display  to  9.9  LPM.  Press  and  hold  the  [

▼]  % 

OXYGEN key. The % OXYGEN display should decrease to 
30%. The audible alarm sounds at the lower limit while the 
key  is  held.  Press  and  hold  the  [

▲]  %  OXYGEN  key,  as 

shown in Figure 1. The % OXYGEN display should increase 
to 100%. An audible alarm sounds at the upper  limit while 
the key is held. 

DIRECTIONS FOR USE

 

CAUTION: 

U s e   p a t ie n t   o b s e r va t io n   t o  

p r e v e n t   o ve r   s e d a t io n   i n   t h e   e ve n t   o f   a n  
o x yg e n   f a i ls a f e   m a lf u n c t i o n   o r   c r o s s e d  

l in e s .   I f   a   p a t ie n t   b e c o m e s   o v e r   s e d a t e d   w h e n  
b e i n g  d e l i v e r e d  1 0 0 %  o x yg e n  [ d u r i n g  a n  a p p a r e n t  
l o s s   o f   n it r o u s   o x i d e   s u p p l y] ,   i t   is   a   d e f i n it e  
i n d ic a t i o n  

o f  

c r o s s e d  

l i n e s  

o r  

a  

f a i ls a f e  

m a lf u n c t i o n .   R e m o v e   t h e   m a s k   im m e d ia t e l y  a n d  
e n c o u r a g e   m o u t h   b r e a t h i n g .   D e l i v e r   p u r e   o x yg e n  
f r o m   a n   o x yg e n   d e m a n d   v a l ve   o n l y  i f   t h e   o x yg e n  
s o u r c e   is   in d e p e n d e n t   f r o m   t h e   s u s p e c t e d  
c r o s s e d   l i n e s   a r e a .  

Note:

 

These  directions  detail  a  basic  delivery  technique. 

However, this is not a comprehensive description and not a 
substitute for a training course that emphasizes a practical, 
hands-on  approach  together  with  instruction  on  safe 
administration techniques. 

Topics covered in such a course 

will  include  experiences  of  practitioners  in  specific  clinical 
settings,  the  pharmacokinetic  properties  of  Nitrous  Oxide, 

strategies  to  avoid  over  sedation  and  allow  for  biological 
variability, and strategies to maximize patient satisfaction.  

Important:  Before  using  the  Digital  MDM  mixer,  check  the 
operation of the unit by performing the Preparation for Use 
Tests  to  ensure  that  your  analgesia  unit  is  performing 
correctly. 

Basic Delivery Technique: 

Practice  titration.  Titration  is  a  method  of  administering  a 
substance  by  adding  definitive  amounts  of  a  drug  until  an 
endpoint  is  reached.    For  nitrous  oxide/oxygen  (N

2

O/O

2

sedation, N

2

O is given in incremental doses until a patient 

has  reached  a  comfortable  relaxed  state  of  sedation.  The 
ability  to  titrate  N

2

O  is  a  significant  advantage  because  it 

limits  the  amount  of  drug  to  that  which  is  required  by  the 
patient.  If  titration  is  done  properly,  the  patient  does  not 
receive any more of the drug than is necessary. The amount 
of N

2

O required by a patient on any given day or time varies. 

 

 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...