background image

 

 

12

 

10545100-Australia Rev. 

– 08/2018 

Directions: 

1.  Open N

2

O / O

2

 cylinder valves 

2.  Press the On/Off key to On state. 

3.  Using the LPM FLOW key pad, set flow rate of O

2

 to desired rate, keep bag about 3/4 full. 

4.  Set % O

2

 concentration to desired level as read by percentages shown on the digital display. Use arrow controls until 

desired  level  is  achieved.  Practice  titration  with  10%  nitrous  upward  movements  every  60  seconds  until  endpoint 
achieved.  Patients  may  typically  experience  relief  of  anxiety,  tingling  in  extremities,  and  euphoria.  Patients  typically 
require less than 50% Nitrous. 

5.  Watching  primary  digital  display,  LPM  FLOW  may  be  re-adjusted  to  bring  the  total  flow  of  gases  to  desired  level. 

Breathing bag will ideally go from ¼ to ¾ full during inhalation and exhalation cycles. 

6.  When the procedure is nearing completion, amounts of N

2

O should be decreased. Terminate the flow of N

2

O and deliver 

100% O

2

 to begin a minimum post-oxygenation period of 3 to 5 minutes. Assess the patient for appropriate recovery. 

Administer  additional  O

2

  if  necessary.  Titration  and  post-procedure  100%  O

2

  will  minimize  Nitrous  exposure  to  the 

operatory, potential patient side effects of lethargy, headache, or nausea, and any potential adverse effects of Nitrous 
diffusion into air filled cavities. 

7.  When procedure is finally completed, turn off using On/Off key. 

8.  Turn OFF the gas supply of the cylinders at the end of the day. 

MAINTENANCE

The Digital MDM Mixer is a precision analgesia mixing unit. 
As  with  any  medical  instrument,  it  requires  periodic 
inspection  and  maintenance.  The  following  routine 
maintenance  procedures  should  be  performed  to  ensure 
system integrity and trouble-free service. 

 

Inspect  mixer,  hoses,  and  connections  daily  for 
damage, wear, and leaks. 

 

Perform functional tests prior to each use. These tests 
are described in 

the “Preparation for Use” section. 

 

While cleaning the outside of the unit, observe caution 
so  that  liquid  does  not  get  inside.  Matrx  by  Parker 
recommends using  a damp (wrung toward dry) cloth 
moistened with a mild soap or a plastic-safe cleaner. 
Do  not  spray  directly  on  the  case.  A  hard  surface 
disinfectant may be applied in the same manner.  

 

 

CAUTION: 

DO NOT submerge the unit

—use a 

damp cloth. 

 

 

 

WARNING: 

Before 

cleaning 

the 

unit, 

disconnect  the  DIGITAL  MDM  mixer’s  power 
cord from the electrical power supply. 

Fuse Replacement 220-240V 

1.  To replace a blown fuse, turn the unit OFF, disconnect 

power to the unit by unplugging the power cord from the 
wall,  and  then  remove  the  power  cord  from  the  input 
connector on the rear of the unit. 

2.  Open the fuse compartment, as shown in Figure 5.  

3.  Remove each blown fuse from its clip and replace it with 

a 5mm x 20mm cartridge fuse, rated 2.5 amp, 250 volt, 
slo-blo (time delay), part number 63882537. Two fuses 
are required for the unit to operate. 

 

Figure 5:

 Fuse Replacement 

 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Reviews: