GENERAL INSTALLATION

Note

: Refer to FM-

916 Mobile “E” Cylinder Stand Installation 

and Instructions for connecting gas supply to the flowmeter 
and cylinder installation to mobile stand. 

Installation of Rubber Goods (Figure 2) 

1.  Slide  the  breathing  bag  opening  over  the  outside 

diameter of the bag mount. 

2.  Slide the 22mm fresh gas connector from the scavenger 

into the breathing tube connector. 

3.  If an accessory, such as the Scavenger Control Valve is 

used, connect it following the instructions supplied with 
it. 

 

Figure 2: 

Installation of Rubber Goods.

 

Connection of Oxygen 

– Nitrous Oxide 

For  safety  reasons,  gas  connections  are  color-coded  and 
mechanically  indexed  using  the  Sleeve  Indexed  System 
(SIS)  to  prevent  accidental  cross  connection.  SIS  female 
hexagon fittings are standard for both hoses. 

1.  Connect  the  SIS  female  hexagon  fitting  of  the  nitrous 

oxide  hose  to  the  male  SIS,  N

2

O  INLET,  fitting  on  the 

back of the mixer head (see Figure 3). Hand-tighten and 
then finish tightening with a 7/8” inch open end wrench 
(approximately 1/8 turn). 

Do not over-tighten

2.  Connect  the  SIS  female  hexagon  fitting  of  the  oxygen 

hose to the male SIS, O

2

 INLET, fitting on the rear of the 

unit  (see  Figure  3).  Hand-tighten  and  then  finish 
tightening  with  a  11/16”  inch  open  end  wrench 
(approximately 1/8 turn). 

Do not over-tighten.

 

Hoses 

Hoses are manufactured for connection to the gas source in 
various configurations: 

Quick  Connect  fittings  are  designed  for  rapid  “no  tool” 
connections  with  corresponding  male  /  female  fittings  or 
outlet  stations.  Refer  to  the  individual  manufacturer’s 
instructions for specific connection procedure. 

SIS  to  SIS  hoses  are  designed  primarily  for  portable 
systems. The gas source (male) SIS fittings are located on 
the regulators on the cylinder mount block provided with the 
portable  units.  Connect  the  DIGITAL  MDM®  mixer  to  the 
cylinder  mount  block  using  the  appropriate  nitrous  oxide 
hose and the oxygen hose. 

 

Figure 3: 

Connection to the back of unit. 

Installation of Power Cord 

Insert the power cord into the input connector on the rear of 
the unit (see Figure 4). 

 

WARNING: 

For  safety  a  Protective  Earth 

(ground)  connection  must  be  made  through 
power cord 3rd wire or Ground Lug (see Fig. 4). 

 

 

 

WARNING: 

Not  providing  a  valid  Protective 

Earth (ground) connection can result in the risk 
of electric shock. 

 

Figure 4: 

Installation of Power Cord.

 

 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...