DEVELOPMENT OF DELIVERY PROTOCOLS 

 

It is the responsibility of the medical establishment and the medical professional to develop and establish specific delivery 
protocols. 
 
The Digital MDM Mixer is designed to deliver a variable concentration of nitrous oxide and oxygen on the continuous flow 
principle. The medical professional will turn on and observe the operating indicators of the device. With the face mask/nasal 
hood sealed every time the patient takes a breath, the patient will receive the full effect of the mixed gas through the face 
mask/nasal hood. 
 
Common procedures conducted with Nitrous Oxide 

– Oxygen Analgesia Conscious Sedation. Ideal for minimally invasive 

procedures where there is a desire for pain management to reduce anxiety or to deliver minimal sedation. 
 
The Digital MDM Mixer is considered transient (less than 60 minutes) in terms of continuous use when providing Analgesia 
(pain management), or Minimal Sedation.  However, a procedure or medical condition that occurs over the course of many 
hours is also considered to be using transient delivery. The upper limit of the number of hours of this described transient 
delivery is at the determination of the medical professional. 
 
Patient Population (Adult and Pediatric):  Used to deliver a gas mixture to a conscious spontaneously breathing patient who 
is awake, alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision 
of  a  healthcare  professional.    Age/Weight  limitations:  see  Relative  Contraindications  (at  the  determination  of  medical 
professional). 
 
 

Note: 

If using a full face mask, Porter recommends using one (DEHP-free and Latex-free) that is biocompatible for medical 

use.  Follow  manufacturer’s  instructions.  The  disposable  mask  materials  have  been  chosen  by  the  medical  device 
manufacturers of the masks intended for medical usage.   Many establishment protocols also call for the use of a bacterial 
filter. Follow manufa

cturer’s instructions. 

 
 

Continuous Flow 

 

 
The Digital MDM Mixer is a continuous flow device. Specific delivery protocols will be established for equipment. 
Continuous flow devices allow the administration of variable concentrations of nitrous oxide and variable flow rates of 
delivered mixture.  

 

Some establishments have developed specific protocols for delivery with continuous flow devices, where the medical 
professional is able to adjust to deliver various percentage mixes, where, with supervision, the delivery may be described 
as “self-administration,” in that the patient will hold the full face mask to the face. Do not use a strap with the full face 
mask. If the patient becomes drowsy the seal around the mask is lost and they will no longer inhale the Gas. It is essential 
that only the patient holds the mask to prevent the onset of deeper stages of analgesia and sedation and loss of protection 
of the laryngeal reflex. 

 

 

 

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Reviews: