SAFETY ALARM SYSTEM

 

There are four alarm conditions on the Digital MDM Mixer. 
They appear in the safety alarm display (see Figure 1, item 
2)  only  when  conditions  warrant.  Otherwise,  they  are  not 
visible. The alarm conditions are:

 

OXYGEN  FAILURE:

  This  alarm  indicates  that  the  set 

oxygen  (O

2

)  flow  cannot  be  delivered.  Check  the  oxygen 

supply.  During  an  OXYGEN  FAILURE  alarm,  the  nitrous 
oxide delivery is automatically shut off. The LPM FLOW and 
% OXYGEN displays are off.

 

NITROUS  FAILURE:

  This  alarm  indicates  that  the  set 

nitrous  oxide  (N

2

O)  flow  cannot  be  delivered.  Check  the 

nitrous  oxide  supply.  During  a  NITROUS  FAILURE  alarm, 
the  LPM  FLOW  display  indicates  oxygen  flow  and  the  % 
OXYGEN display is off. 

Both the OXYGEN FAILURE and NITROUS FAILURE are 
self-canceling  alarms.  When  the  gas  is  restored,  the  unit 
returns to the flow rate and mixture ratio which were in effect 
before the alarm condition. 

1  LPM  OXYGEN  LIMIT:

  This  message  is  displayed  when 

the oxygen flow rate reaches a value of 1 LPM; the unit will 
not allow the oxygen flow rate to go below this value. If the 
user attempts to lower the % OXYGEN while this message 
is  displayed,  the  unit  compensates  by  increasing  the  total 
flow rate. 

FLUSH  FLOW  LOW:

  This  message  indicates  that  the 

oxygen  flush  flow  rate  is  less  than  10  LPM.  Check  the 
oxygen supply pressure.

 

KEY PAD CONTROLS 

 

CAUTION: 

Operate  key  pad  with  finger 

pressure only. Do not use hard or sharp objects. 
Failure  to  do  so  may  result  in  damage  to  the 
front panel. 

On/Off Key:

 The ON/OFF key controls the operation of the 

unit.  Momentarily  depressing  the  ON/OFF  key  changes 
power state from OFF to ON or ON to OFF. 

Note:

  If  the  power  to  the  unit  is  interrupted  and  then 

restored, either by  disconnecting / reconnecting the power 
cord or by local power failure, the unit will display OFF. The 
operator must press the ON/OFF key to restart the unit. The 
unit  will  them  display  default  values  of  100%  oxygen  at  5 
LPM. 

 

CAUTION: 

After turning the device off, or when 

the  unit  is  not  in  use,  all  oxygen  and  nitrous 
oxide supply cylinder valves must be closed. 

Flow Control:

 The LPM FLOW key pad [

▲ or ▼] controls 

the  total  flow  rate  delivered  by  the  unit.  Pressing  the 
appropriate  key  momentarily  increases  or  decreases  the 
total flow rate by 0.1 LPM. By pressing and holding a key, 
the total flow rate indication changes slowly for the first two 
steps, then changes at a faster rate for the remainder of the 
time the key is held. An audible signal indicates that a limit 
has been reached. 

%O

2

 Control:

 The % OXYGEN key pad [

▲ or ▼] controls 

the  percentage  of  oxygen  delivered  by  the  unit.  Press  the 
appropriate  key  momentarily  to  increase  or  decrease  the 
oxygen  percentage  one  step  (1%  or  5%).  Pressing  and 
holding the key changes the oxygen percentage slowly for 
the first two steps, then faster for the remainder of the time 
the  key  is  held.  This  varies  the  ratio  of  oxygen  to  nitrous 
oxide.  An  audible  signal  indicates  that  a  limit  has  been 
reached. 

 

 

 

Setting % OXYGEN Step Amount (1% or 5%):

 The Digital 

MDM  Mixer  is  shipped  from  the  factory  programmed  to 
change 1% for each press of the % OXYGEN key pad [

▲ or 

▼].  The  amount  can  be  changed  from  1%  to  5%  by 
performing the following: 

1.  Press and hold the DISPLAY N

2

O key. 

2.  While pressing the DISPLAY N

2

O key, press the 

DISPLAY O

2

 key until the audible alarm sounds. 

3.  Release both keys. 

The audible signal indicates the change has occurred and is 
programmed  in  memory.  The  unit  will  remain  in  the  new 
mode until this sequence is repeated. Unplugging or turning 
the unit off will not affect the programmed mode. 

Flow  Display  Control:

  The  key  pad  DISPLAY  O

2

  or 

DISPLAY N

2

O select the gas flow rate displayed on the flow 

display. Press and hold the DISPLAY N

2

O key to display the 

nitrous oxide component of the total flow. Press and hold the 
DISPLAY  O

2

  key  to  display  the  oxygen  component  of  the 

total  flow.  If  neither  key  is  held,  the  LPM  FLOW  display 
indicates the total gas flow. 

Oxygen  Flush  Key:

  Pushing  the  O

2

  FLUSH  key  delivers 

100% oxygen directly to the mixed gas outlet at the rear of 
the unit. This outlet is connected to the patient through the 
breathing  tube  connector  (see  Figure  2).  While  the  key  is 
pushed,  100%  oxygen  is  delivered  continuously  at  an 
approximate flow rate of 20 LPM. Releasing the key returns 
the unit to the flow rate and mixture ratio which were in effect 
prior to the flush. 

Note

: Whenever the unit is turned ON an automatic 100% 

oxygen ‘flush’ will be briefly delivered to the patient. 

Alarm Test/Silence Key:

 The alarm test / silence key 

serves two functions. When pressed, all the display, 
indicators and the audible alarm turns ON. This key also 
silences the audible alarm when a NITROUS FAILURE 
condition exists. The audible alarm will be silent for about 
five minutes and then resume, unless the failure condition 
has been corrected

Summary of Contents for Matrx Digital MDM

Page 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Page 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Page 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Page 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Page 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Page 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Page 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Page 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Page 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Page 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Page 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Page 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Page 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Reviews: