ESSENTIAL PERFORMANCE
There are no essential performance characteristics according to the risk assessment.
ELECTROMAGNETIC ENVIRONMENT
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such environment.
EMISSION TEST
Emission test
Compliance
Electromagnetic environment — Guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its
RF-emission is very low and not likely to cause any interference nearby
electronic equipment.
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations /flicker
emissions
IEC 61000-3-3
Passed
IMMUNITY TEST
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment —
Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
Contact: ±8 kV
Air: ±15 kV
Contact: ±8 kV
Air: ±15 kV
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient /burst
IEC 61000-4-4
5/50 ns,
100 kHz, ±2 kV
5/50 ns,
100 kHz, ±2 kV
Mains power quality should
be similar to that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
1.2/50 (8/20) µs
LtL: ±1 kV
LtG: ±2 kV
1.2/50 (8/20) µs
LtL: ±1 kV
LtG: ±2 kV
Mains power quality should
be similar to that of a typical
commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0 % UT for 0.5 cycle (1 phase)
0 % UT for 1 cycle
70 % UT for 25/30 cycles (50/60
Hz)
0 % UT for 250/300 cycles (50/60
Hz)
0 % UT for 0.5 cycle (1 phase)
0 % UT for 1 cycle
70 % UT for 25/30 cycles (50/60
Hz)
0 % UT for 250/300 cycles (50/60
Hz)
Mains power quality should
be similar to that of a typical
commercial or hospital
environment. When the user of the
device requires continuous function
in the event of disruption of supply,
it is recommended the device is
operated from an uninterruptible
power supply or battery.