G.  MAINTENANCE OF YOUR 

The filter on the front of the PRONEB

®

 

max

 Compressor 

was designed to be monitored without removing it. This 
makes it easy to check without much work. When the 
filter is new, it is bright white in color. With use, the 
filter will turn grey, indicating the need for replacement. 
The filter must be replaced immediately if it gets wet or 
becomes clogged. A good rule to follow is to replace 
your filter every 6 months - at the same time as your 
PARI reusable nebulizer.
For filter replacement, obtain the replacement filter 
set part #130F1002P2 which includes the filter and 
filter cap (Fig E). Pull the old filter and cap out of the 
compressor and discard (Fig F).
Once the old filter and cap are removed, tilt the 
compressor so that the face of the unit is facing the 
floor. Gently tap on the back of the compressor to 
remove any unwanted materials from the filter area. 
Place the compressor on a stable surface and using 
clean hands, push a new filter with cap into the filter 
receptacle.

NOTE: 

DO NOT ATTEMPT TO WASH AND 

REUSE AIR FILTERS. DO NOT USE SUBSTITUTE 
FILTER MATERIAL AS COMPRESSOR DAMAGE 
MAY RESULT.

COMPRESSOR

Use a damp cloth to wipe the exterior of your PRONEB

®

 

max

 housing and tubing.

CAUTION

Never spray fluid into the ventilation slots on your 
PRONEB

®

 

max

. Fluid could cause damage to the electrical 

& compressor parts and could lead to malfunction. 
In the event fluid enters the unit, please contact PARI 
Patient Connection at 1.800.FAST.NEB (327.8632).

MAINTENANCE & REPAIR

Never operate your PRONEB

®

 

max

 if it is improperly or 

incompletely assembled or damaged. Use only original 
PARI parts and attachments authorized by PARI for the 
PRONEB

®

 

max

 Aerosol Delivery System. Use of other 

parts and attachments may damage the unit or cause 
improper treatment. Do not open or disassemble the 
compressor unit at any time.

NOTE: 

THERE ARE NO SERVICEABLE PARTS 

INSIDE THE PRONEB

®

 MAX AEROSOL DELIVERY 

SYSTEM. SERVICE BY ANYONE VOIDS 
WARRANTY.

E

F

PRONEB

®

 MAX

PRONEB

®

 

max

 REPLACEMENT ITEMS

Part No.

LC Sprint

®

 Reusable Nebulizer 

023F35

LC PLUS

®

 Reusable Nebulizer

022F81

Wing Tip

®

 Tubing

044E0050

PRONEB

®

 

max

 Replacement Filter 2pk

130F1002P2

Summary of Contents for PRONEB

Page 1: ...SAFETY PRECAUTIONS BEFORE USE READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE USE THE PRODUCT ONLY FOR ITS INTENDED USE AS DESCRIBED IN THESE INSTRUCTIONS FOR USE IMPROPER USE CAN RESULT IN SERIOUS OR...

Page 2: ...en prescribed by a physician and with the prescribed medication In order to achieve the specified droplet characteristics the compressor or central air supply must provide 3 5 8 lpm at a minimum of 20...

Page 3: ...is plug will fit in a polarized outlet only one way If the plug does not fit fully in the outlet reverse the plug 5 Never block the air vents of this compressor or place it on a soft surface such as a...

Page 4: ...le Nebulizer on hand at all times Air Outlet Air Filter PARI Reusable Nebulizer Holder Power Switch Wing Tip Tubing PARI Reusable Nebulizer 2 included Inspiratory Valve Cap Nebulizer Insert Nebulizer...

Page 5: ...izer outlet with the expiratory valve facing up Then close the inspiratory valve cap by pressing it firmly onto the top of the nebulizer If using a mask attach mask to the nebulizer outlet and refer t...

Page 6: ...treatment day soak all parts of the PARI Reusable Nebulizer except tubing and mask in a solution of 1 part distilled white vinegar and 3 parts hot tap water for 1 hour You can substitute respiratory...

Page 7: ...ds push a new filter with cap into the filter receptacle NOTE DO NOT ATTEMPT TO WASH AND REUSE AIR FILTERS DO NOT USE SUBSTITUTE FILTER MATERIAL AS COMPRESSOR DAMAGE MAY RESULT COMPRESSOR Use a damp c...

Page 8: ...ssure 700 hpa to 1060 hpa EXPLANATION OF SYMBOLS Attention Read all Instructions for Use before operation Insulation enclosed Class II equipment Type BF Protection from electrical shock Do not dispose...

Page 9: ...el Electromagnetic environment Guidance Electrostatic discharge ESD IEC 61000 4 2 Contact 8 kV Air 15 kV Contact 8 kV Air 15 kV Floors should be wood concrete or ceramic tile If floors are covered wit...

Page 10: ...electromagnetic field Amplitude modulated IEC 61000 4 3 80 MHz 2 7 GHz 10 V m T Home Healthcare T Prof Healthcare T 80 1 kHz WARNING Portable RF communications equipment including peripherals such as...

Page 11: ...NG IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO SIXTY 60 MONTHS FROM THE DATE OF FIRST RETAIL PURCHASE ALL EXPRESS OR IMPLIED WARRANTIES ON ANY OTHER PARTS...

Page 12: ...S ADVERTENCIAS E INSTRUCCIONES ANTES DE UTILIZAR EL EQUIPO USE EL PRODUCTO SOLO PARA EL PROP SITO QUE SE DESCRIBE EN ESTAS INSTRUCCIONES DE USO UN USO INCORRECTO PUEDE PROVOCAR TRASTORNOS LESIONES GRA...

Page 13: ...camente por indicaci n m dica y con el medicamento recetado Para lograr las caracter sticas de gota especificadas el compresor o suministro de aire central debe proporcionar de 3 5 a 8 lpm a un m nimo...

Page 14: ...e en una toma polarizada una sola posici n Si el enchufe no entra por completo en la toma delo vuelta 5 Nunca bloquee las ranuras de ventilaci n del compresor y tampoco ponga el compresor sobre superf...

Page 15: ...spiratoria Inserto del nebulizador Semic rculo del nebulizador Anillos de alineaci n Taza del nebulizador Entrada de aire Boquilla con v lvula espiratoria Salida del nebulizador Ubicaci n sugerida par...

Page 16: ...a v lvula inspiratoria presion ndola firmemente sobre la parte superior del nebulizador Si usa una mascarilla inst lela en la salida del nebulizador y consulte las instrucciones de la mascarilla D PRE...

Page 17: ...oluci n de una parte de vinagre blanco destilado y 3 partes de agua caliente del grifo durante 1 hora Puede reemplazar el desinfectante de equipos respiratorios como Control III por vinagre blanco des...

Page 18: ...del filtro NOTA NO INTENTE LAVAR Y REUTILIZAR LOS FILTROS DE AIRE NO USE OTRO MATERIAL PARA REEMPLAZAR EL FILTRO PUESTO QUE PUEDE DA AR EL COMPRESOR COMPRESOR Use un pa o h medo para limpiar el exteri...

Page 19: ...a 700 hPa a 1060 hPa EXPLICACI N DE S MBOLOS Atenci n Lea todas las Instrucciones de uso antes de utilizar el sistema Equipo de aislamiento cerrado Clase II Protecci n tipo B contra descarga el ctrica...

Page 20: ...4 2 Contacto 8 kV Aire 15 kV Contacto 8 kV Aire 15 kV El suelo debe ser de madera cemento o baldosas de cer mica Si el suelo est revestido de material sint tico la humedad relativa debe ser por lo men...

Page 21: ...modulada IEC 61000 4 3 80 MHz 2 7 GHz 10 V m T Atenci n m dica dom stica T Atenci n m dica profesional T 80 1 kHz ADVERTENCIA Los equipos de comunicaciones por RF port tiles incluidos perif ricos como...

Page 22: ...LIDAD Y ADECUACI N PARA UN PROP SITO PARTICULAR SE LIMITAN A SESENTA 60 MESES A PARTIR DE LA FECHA DE LA PRIMERA COMPRA EN EL MERCADO MINORISTA TODAS LAS GARANT AS EXPRESAS O IMPL CITAS SOBRE CUALQUIE...

Reviews: