INDICATIONS FOR USE: PRONEB

®

 MAX AEROSOL DELIVERY SYSTEM

The PARI PRONEB

®

 

max

 is a tabletop, AC-powered air compressor intended to provide a source of compressed air for 

use with jet nebulizers with patients for whom doctors have prescribed medication for nebulization. It is intended for 
adult and pediatric patients 2 years and older, and may be used in hospitals, hospital-type facilities, nursing homes, 
sub-acute institutions and home environments.

WARNING 

Use only with nebulizers and tubing authorized by PARI 

for this device. Other nebulizers and tubing may result in 

improper treatment.
When using electrical and medical products, especially 

when children are present, basic safety precautions must 

be followed.

DANGER

To reduce the risk of serious or fatal injury from 

electrocution:

1. 

Do not place or store product where it can fall or be 

pulled into a bathtub, sink, water, or other liquid. Do 

not place in or drop into water or other liquid. Do not 

use while bathing.

2. 

Do not reach for a product that has fallen into water 

or other liquid. Unplug immediately.

WARNING 

To reduce the risk of serious or fatal injury from 

electrocution, fire or burns and to reduce the risk of 

damage and malfunction to the product:

1. 

Before use, check the serial number label on the 

bottom of the PARI PRONEB

®

 

max

 Compressor to 

ensure that the voltage and current indicated on 

the compressor label correspond to the voltage and 

current available.

2. 

Never operate this product if it has a damaged cord 

or plug, if it is not working properly, if it has been 

dropped or damaged, or if it has been exposed to any 

liquids inside the case. Return the product to PARI for 

examination and repair.

3. 

Keep all electrical cords away from heated surfaces. Do 

not overload wall outlets or use extension cords.

4. 

This appliance has a polarized plug (one blade is wider 

than the other). As a safety feature, this plug will fit in 

a polarized outlet only one way. If the plug does not fit 

fully in the outlet, reverse the plug.

5. 

Never block the air vents of this compressor or place it 

on a soft surface, such as a bed or couch, where the 

air vents may be blocked. Keep the air vents free of 

lint, hair, and similar substances. Blocked vents could 

cause excessive heat buildup. This product should 

never be left unattended when plugged in.

6. 

Do not operate outdoors.

7. 

Not suitable for use in the presence of flammable 

anesthetic mixture with air or oxygen or nitrous oxide.

8. 

Do not operate where oxygen is being administered in 

a closed environment such as an oxygen tent.

9. 

Never drop or insert any object into any openings.

10. 

Always unplug this product immediately after using.

11. 

Be sure the PARI PRONEB

®

 

max

 Compressor is 

unplugged prior to cleaning.

12. 

Do not open or disassemble the PARI PRONEB

®

 

max

 

Compressor. Refer all servicing to PARI.

WARNING 

To reduce risk that the compressor will fail to operate 

properly as a result of decreased electromagnetic 

interference, resulting in improper treatment:

1. 

Portable and mobile radio frequency (“RF”) 

communication devices can disrupt electrical  

medical equipment.

2. 

Use only with accessories and cables authorized 

by PARI for this device. Use of this device with 

unauthorized accessories and cables may result in 

increased electromagnetic emissions or decreased 

electromagnetic immunity of the device.

3. 

This device should not be used adjacent to, or stacked 

with other equipment. If adjacent or stacked use is 

necessary, this device should be observed to verify 

normal operation.

B. SAFETY PRECAUTIONS–PRONEB

®

 MAX AEROSOL DELIVERY SYSTEM

ENVIRONMENT OF USE:

As a tabletop compressor, it is intended for use in home healthcare and professional care settings. It is not permitted for 

use in trains, motor vehicles, or airplanes.
Within professional care settings, the device must only be used in in-patient wards and intensive care units. The device 

must never be used in areas exposed to elevated magnetic or electrical radiation (e.g. in the vicinity of an MRI scanner or 

electromagnetic security equipment).

Summary of Contents for PRONEB

Page 1: ...SAFETY PRECAUTIONS BEFORE USE READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE USE THE PRODUCT ONLY FOR ITS INTENDED USE AS DESCRIBED IN THESE INSTRUCTIONS FOR USE IMPROPER USE CAN RESULT IN SERIOUS OR...

Page 2: ...en prescribed by a physician and with the prescribed medication In order to achieve the specified droplet characteristics the compressor or central air supply must provide 3 5 8 lpm at a minimum of 20...

Page 3: ...is plug will fit in a polarized outlet only one way If the plug does not fit fully in the outlet reverse the plug 5 Never block the air vents of this compressor or place it on a soft surface such as a...

Page 4: ...le Nebulizer on hand at all times Air Outlet Air Filter PARI Reusable Nebulizer Holder Power Switch Wing Tip Tubing PARI Reusable Nebulizer 2 included Inspiratory Valve Cap Nebulizer Insert Nebulizer...

Page 5: ...izer outlet with the expiratory valve facing up Then close the inspiratory valve cap by pressing it firmly onto the top of the nebulizer If using a mask attach mask to the nebulizer outlet and refer t...

Page 6: ...treatment day soak all parts of the PARI Reusable Nebulizer except tubing and mask in a solution of 1 part distilled white vinegar and 3 parts hot tap water for 1 hour You can substitute respiratory...

Page 7: ...ds push a new filter with cap into the filter receptacle NOTE DO NOT ATTEMPT TO WASH AND REUSE AIR FILTERS DO NOT USE SUBSTITUTE FILTER MATERIAL AS COMPRESSOR DAMAGE MAY RESULT COMPRESSOR Use a damp c...

Page 8: ...ssure 700 hpa to 1060 hpa EXPLANATION OF SYMBOLS Attention Read all Instructions for Use before operation Insulation enclosed Class II equipment Type BF Protection from electrical shock Do not dispose...

Page 9: ...el Electromagnetic environment Guidance Electrostatic discharge ESD IEC 61000 4 2 Contact 8 kV Air 15 kV Contact 8 kV Air 15 kV Floors should be wood concrete or ceramic tile If floors are covered wit...

Page 10: ...electromagnetic field Amplitude modulated IEC 61000 4 3 80 MHz 2 7 GHz 10 V m T Home Healthcare T Prof Healthcare T 80 1 kHz WARNING Portable RF communications equipment including peripherals such as...

Page 11: ...NG IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO SIXTY 60 MONTHS FROM THE DATE OF FIRST RETAIL PURCHASE ALL EXPRESS OR IMPLIED WARRANTIES ON ANY OTHER PARTS...

Page 12: ...S ADVERTENCIAS E INSTRUCCIONES ANTES DE UTILIZAR EL EQUIPO USE EL PRODUCTO SOLO PARA EL PROP SITO QUE SE DESCRIBE EN ESTAS INSTRUCCIONES DE USO UN USO INCORRECTO PUEDE PROVOCAR TRASTORNOS LESIONES GRA...

Page 13: ...camente por indicaci n m dica y con el medicamento recetado Para lograr las caracter sticas de gota especificadas el compresor o suministro de aire central debe proporcionar de 3 5 a 8 lpm a un m nimo...

Page 14: ...e en una toma polarizada una sola posici n Si el enchufe no entra por completo en la toma delo vuelta 5 Nunca bloquee las ranuras de ventilaci n del compresor y tampoco ponga el compresor sobre superf...

Page 15: ...spiratoria Inserto del nebulizador Semic rculo del nebulizador Anillos de alineaci n Taza del nebulizador Entrada de aire Boquilla con v lvula espiratoria Salida del nebulizador Ubicaci n sugerida par...

Page 16: ...a v lvula inspiratoria presion ndola firmemente sobre la parte superior del nebulizador Si usa una mascarilla inst lela en la salida del nebulizador y consulte las instrucciones de la mascarilla D PRE...

Page 17: ...oluci n de una parte de vinagre blanco destilado y 3 partes de agua caliente del grifo durante 1 hora Puede reemplazar el desinfectante de equipos respiratorios como Control III por vinagre blanco des...

Page 18: ...del filtro NOTA NO INTENTE LAVAR Y REUTILIZAR LOS FILTROS DE AIRE NO USE OTRO MATERIAL PARA REEMPLAZAR EL FILTRO PUESTO QUE PUEDE DA AR EL COMPRESOR COMPRESOR Use un pa o h medo para limpiar el exteri...

Page 19: ...a 700 hPa a 1060 hPa EXPLICACI N DE S MBOLOS Atenci n Lea todas las Instrucciones de uso antes de utilizar el sistema Equipo de aislamiento cerrado Clase II Protecci n tipo B contra descarga el ctrica...

Page 20: ...4 2 Contacto 8 kV Aire 15 kV Contacto 8 kV Aire 15 kV El suelo debe ser de madera cemento o baldosas de cer mica Si el suelo est revestido de material sint tico la humedad relativa debe ser por lo men...

Page 21: ...modulada IEC 61000 4 3 80 MHz 2 7 GHz 10 V m T Atenci n m dica dom stica T Atenci n m dica profesional T 80 1 kHz ADVERTENCIA Los equipos de comunicaciones por RF port tiles incluidos perif ricos como...

Page 22: ...LIDAD Y ADECUACI N PARA UN PROP SITO PARTICULAR SE LIMITAN A SESENTA 60 MESES A PARTIR DE LA FECHA DE LA PRIMERA COMPRA EN EL MERCADO MINORISTA TODAS LAS GARANT AS EXPRESAS O IMPL CITAS SOBRE CUALQUIE...

Reviews: