Questions? 1.800.FAST.NEB (327.8632) | PARI.com/PRONEBMAX
EN-9
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment —
Guidance
Power frequency (50 Hz / 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Amplitude modulated
150 kHz – 80 MHz
3 V
ISM and amateur radio bands
6 V
T 80 % / 1 kHz
150 kHz – 80 MHz
3 V
ISM and amateur radio bands
6 V
T 80 % / 1 kHz
WARNING: Portable RF
communications equipment
(including peripherals such as
antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part
of the device, including cables,
specified by the manufacturer.
Otherwise, degradation of the
performance of this equipment
could result.
Radio-frequency
electromagnetic field
Amplitude modulated
IEC 61000-4-3
80 MHz – 2.7 GHz
10 V/m
T Home Healthcare
T Prof. Healthcare
T 80 % / 1 kHz
WARNING: Portable RF
communications equipment
(including peripherals such as
antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part
of the device, including cables,
specified by the manufacturer.
Otherwise, degradation of the
performance of this equipment
could result.
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
380 – 390 MHz 27 V/m;
PM 50 %;
18 Hz
430 – 470 MHz 28 V/m;
(FM ±5 kHz, 1 kHz sine) PM;
18 Hz
704 – 787 MHz 9 V/m;
PM 50 %;
217 Hz
800 – 960 MHz 28 V/m;
PM 50 %;
18 Hz
1700 – 1990 MHz 28 V/m;
PM 50 %;
217 Hz
2400 – 2570 MHz 28 V/m;
PM 50 %;
217 Hz
5100 – 5800 MHz 9 V/m;
PM 50 %;
217 Hz
WARNING: Portable RF
communications equipment
(including peripherals such as
antenna cables andexternal
antennas) should be used no closer
than 30 cm (12 inches) to any part
of the device, including cables,
specified by the manufacturer.
Otherwise, degradation of the
performance of this equipment
could result.