Intended use
The eFlow
®
rapid
inhalation system is
intended for the treatment of diseases of
the airways and lungs.
Proper use
The nebuliser system must only be oper-
ated in accordance with its intended pur-
pose. The frequency of application may
vary depending on the clinical picture and
the medication used. The nebuliser hand-
set is designed for use by one patient only.
Only original PARI accessories must be
used to operate the system.
PARI Pharma GmbH is not responsible
for damage if the system is used improp-
erly or contrary to the instructions.
Intended patient groups
The nebuliser system is only suitable for
use by patients who can breathe unas-
sisted and are conscious.
Inhalation with the mouthpiece is the most
effective form of inhalation, since it results
in minimum medication loss on the way to
the lung. For babies and infants who can-
not use the mouthpiece to inhale, PARI
offers appropriate masks as accessories.
Medications
The eFlow
®
rapid
inhalation system must
only be used to nebulise permitted inhala-
tion solutions and suspensions. These are
usually specified by the doctor. Take note
of any limitations in the package leaflet
accompanying each medication.
Life cycle
- Controller: 3 to 5 years
- Nebuliser handset (without aerosol
head): 1 year
- Aerosol head: 3 to 6 months
The expected life cycle of the aerosol
head depends on how often and for how
long it is used. With two inhalation treat-
ments lasting a total of 30 minutes’ inha-
lation time and one disinfection per day,
a life cycle of up to 6 months can be
expected.
Info:
The eBase Controller can also be
used for a medication-specific nebu-
liser handset. Follow the instructions
for use of the nebuliser handset as
well.
CAUTION:
Medications that are approved for inhala-
tion with a medication-specific nebuliser
handset must not be used with an
eFlow
®
rapid
nebuliser handset (risk of
incorrect dosage).
CAUTION:
Using accessories (components or
cables) other than those supplied as spare
parts by the manufacturer may cause
increased electromagnetic emissions or
reduced electromagnetic immunity, lead-
ing to a malfunction.
CAUTION:
Never attempt to inhale essential oils, as
this can cause irritation of the airways
(intense coughing).
Summary of Contents for eFlow Rapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 04 18...
Page 32: ...28 eFlow rapid 04 18...
Page 58: ...54 eFlow rapid 04 18...
Page 108: ...104 eFlow rapid 04 18...
Page 134: ...130 eFlow rapid 04 18...
Page 160: ...156 eFlow rapid 04 18...
Page 186: ...182 eFlow rapid 04 18...
Page 259: ......