The nebulisation period has become
noticeably longer.
If the nebuliser system is used regularly, the
nebulisation period will become longer. This
is not a malfunction. However, if you notice
that the nebulisation period has become con-
siderably longer, this may be due to one of
the following causes:
- You are inhaling a different medication.
- The fill volume has been increased.
- The medication reservoir cap was not
closed tightly.
- The aerosol head was not cleaned immedi-
ately after the last application, or it is visibly
damaged.
- You are using the aerosol head for longer
or more often than is described in section
Life cycle.
If none of the points listed above apply, carry
out the following function test for the aerosol
head:
• Pour 2.5 ml isotonic sodium chloride solu-
tion (0.9%) into the medication reservoir.
• Switch the controller on and note how long
nebulisation takes without inhaling.
If the nebulisation period is longer than
4 minutes, use the
easycare
cleaning aid.
Take note of the information regarding fre-
quency of use of the
easycare
cleaning aid
in the accompanying instructions for use.
• If the nebulisation period is still longer than
5 minutes after cleaning with the
easycare
cleaning aid, you must replace the aerosol
head.
Info:
The time specified above only applies
for 2.5 ml isotonic sodium chloride
solution (0.9%).
Aerosol escapes continuously in large
volumes from the opening slots in the
medication reservoir.
• Check whether the nebuliser handset has
been assembled correctly.
When doing so, check that the two valve
flaps of the inspiratory valve are seated
correctly. They must lie perfectly flush.
Fault
Possible cause/Remedy
Summary of Contents for eFlow Rapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 04 18...
Page 32: ...28 eFlow rapid 04 18...
Page 58: ...54 eFlow rapid 04 18...
Page 108: ...104 eFlow rapid 04 18...
Page 134: ...130 eFlow rapid 04 18...
Page 160: ...156 eFlow rapid 04 18...
Page 186: ...182 eFlow rapid 04 18...
Page 259: ......