42
eFlow
®
rapid
Cleaning
Clean the nebuliser handset (incl. aerosol
head) immediately after each application.
• Place all nebuliser handset parts as well
as the aerosol head in warm tap water
with a little washing up liquid for approx.
5 minutes. If the parts are heavily soiled,
move the nebuliser handset parts gently
back and forth in the washing water.
• Rinse all nebuliser handset parts and
the front and back of the aerosol head
thoroughly under running tap water.
• Shake all parts to remove excess water
more quickly.
• Place the individual parts on a dry, clean
surface and allow to dry completely:
Disinfection
Disinfect the nebuliser handset parts and
the aerosol head at least once a day, e.g.,
immediately after cleaning at the end of
the day (a nebuliser handset cannot be
disinfected effectively unless it has been
cleaned first).
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
each cleaning.
NOTE:
Do not attempt to disinfect the aerosol
head in a microwave oven as it cannot be
ensured that the product components will
not be damaged.
Summary of Contents for eFlow Rapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 04 18...
Page 32: ...28 eFlow rapid 04 18...
Page 58: ...54 eFlow rapid 04 18...
Page 108: ...104 eFlow rapid 04 18...
Page 134: ...130 eFlow rapid 04 18...
Page 160: ...156 eFlow rapid 04 18...
Page 186: ...182 eFlow rapid 04 18...
Page 259: ......