eFlow
®
rapid
43
en
With a commercially available
disinfector
• To ensure effective disinfection, use a
thermal disinfector with an operating
time of at least 6 minutes. With regard to
performing the disinfection, the duration
of the disinfection process and the
quantity of water required, follow the
instructions for use of the disinfector you
are using.
• Take the parts out of the disinfector
immediately after disinfection is com-
pleted. Place them on a dry, clean sur-
face and allow them to dry completely:
• Check the nebuliser handset parts after
every disinfection. Replace all broken,
misshapen and heavily discoloured
parts.
Possible alternative:
Boiling in distilled water
• Put all parts in boiling water for at least
5 minutes. Use a clean saucepan and
distilled water for this.
• Take the parts out of the saucepan.
• Shake all parts to remove excess water
more quickly.
• Place the individual parts on a dry, clean
surface and allow to dry completely:
• Check the nebuliser handset parts after
every disinfection. Replace all broken,
misshapen and heavily discoloured
parts.
CAUTION:
Inadequate disinfection promotes the
growth of bacteria and thus increases
the risk of infection. If disinfection is not
carried out for the time specified for the
device concerned, it will not be effective.
Therefore, do not switch the device off
before the disinfection cycle is com-
plete. Also, make sure that the device is
clean and check it regularly for proper
operation.
NOTE:
Plastic will melt if it comes into contact
with the hot bottom of the saucepan.
Therefore, make sure that there is suffi-
cient water in the pan. This will prevent
the parts from being damaged.
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
every disinfection.
Summary of Contents for eFlow Rapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 04 18...
Page 32: ...28 eFlow rapid 04 18...
Page 58: ...54 eFlow rapid 04 18...
Page 108: ...104 eFlow rapid 04 18...
Page 134: ...130 eFlow rapid 04 18...
Page 160: ...156 eFlow rapid 04 18...
Page 186: ...182 eFlow rapid 04 18...
Page 259: ......