15. Safety-related controls and maintenance
The device must be examined once a year in a technical safety inspection (TSI) and technical measure-
ment evaluation TME - together with this operating manual and the appropriate medical product book.
These inspections must be performed exclusively by persons who can properly execute these examinations
based on their training, expertise and experiences, gained during their practical occupation. They must be
able to perform these inspections without instructions.
Terms: One-year
•
General visual inspection
•
Electrical safety inspection according to VDE 0751 (protective wire inspection, discharge current)
•
Inspection of the safety concept according to valid inspection instructions
•
General functional check
All maintenance must be exclusively performed by companies that have been authorized by PARAVAN me-
dical GmbH to perform maintenance work on the respective device type.