129 von 147
Gener
al
Inf
ormation
Pr
epar
e
Oper
ate
Help
Technology
- Electromagnetic compatibility (EMC) -
mov07_EN_Rev.2.1
Voltage dips, short-
term interruptions
and fluctuations of
the supply voltage
according to 61000-
4-11
< 5 % U
T
(>95 % dip
of U
T
for 0.5 period
40 % U
T
(60% dip of
UT) for 5 periods
70 % U
T
(30% dip of
UT) for 25 periods
< 5 % U
T
(>95 % dip
of U
T
for 5 seconds
Not applicable
The quality of the power voltage
should comply with a typical busi-
ness- or hospital environment.
If the user requires continued
usage of the movanimo even
during disturbances in the ener-
gy supply, it is recommended to
supply the movanimo device from
a disturbance free power supply
or battery.
Magnetic field for
the supply frequency
(50/60 Hz) according
to IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields for the net fre
-
quency should comply with typical
values as they exist in the busi-
ness- and hospital environment.
COMMENT U
T
is the alternating supply voltage prior to application of test levels.
Table 202 according to DIN EN 60601-1-2, 6.8.3.201