54
WARNING
g
Notification is not made during bed operation (e.g., head up, height ad
-
justment) even if the patient gets out of the bed or sits up on the bed. Fur-
thermore, getting into or out of the bed during bed operation may cause
the patient to fall off or fall over, resulting in injury. If you consider that the
patient may get in or out of the bed during bed operation, set Overall lock-
out or Patient lock-out (see p. 65).
g
Before using BED EXIT ALARM, set the notification function properly (for
the setting procedure, see the instruction manual for the relay unit). If the
settings are not correct, detection may not work and notification may not be
made, and the patient may fall off or fall over, resulting in injury. Even when
the settings are correct, detection may not work and notification may not be
made in the following cases.
n
If it takes 4 minutes or more for the patient to get out of the bed or sit up on the bed
n
If the patient gets out of the bed or sits up on the bed after a heavy object (which weighs more than
half the patient’s weight) is placed on the bed
n
If the patient who is lying at the foot end of the bed sits up (SIT-UP function)
n
If the patient sits up while placing his or her hand or elbow on the top of the head end of the bed (SIT-
UP function)
g
When using BED EXIT ALARM, use it at a temperature between 15˚C and
30˚C and humidity between 30% RH and 75% RH. The detection accuracy
may decrease, notification may not be made, and the patient may fall off
or fall over, resulting in injury.
g
Do not place objects such as furniture and fixtures between the bed and
floor. Detection may not work and the patient may fall off or fall over, re
-
sulting in injury.