17
WARNING
g
Do not attach the restraint belt to any location other than the restraint
holder.
The restraint belt may unexpectedly come off resulting in injury, or the bed may be damaged or
deformed.
g
The bed should be transferred by at least two persons.
n
When transferring the bed, blind spots may occur in the direction of travel and the bed may collide
with a person, resulting in injury. In addition, the bed may collide with an object and the bed and the
object may be damaged or deformed.
n
The person at the front should pay attention to the direction of travel and corners.
g
When using suitable products directly attached to the bed such as mat-
tresses, use products suitable for the bed specified by Paramount Bed.
The use of products other than the specified ones or those from other companies may create unintended
gaps, cause contact between products, or lower stability, resulting in injury or failure of the bed. (For
suitable products, see p. 72.)
g
Keep the bed away from fire.
Do not use heating appliances such as heaters near the bed. This may cause deterioration, deformation,
or fire.
g
Do not use the bed in the oxygen tent.
n
Fire could result.
n
When using oxygen administration devices, use nasal type, mask type, or artificial respirator type
only.
g
When using flammable anesthetics, do not operate the bed.
Fire could result.
g
Be careful to prevent clothing from getting caught on suitable products
and protrusions.
n
If the bed is operated with clothing caught, the clothing may be pulled, resulting in injury.
n
When getting on or off the bed, the patient may fall off the bed or fall down, resulting in injury.
1
BEFORE USING THE BED