DigitalFlow™ XGF868i Startup Guide
9
Chapter 1. Installation
1.7.2.1
Local or Remote Electronics Enclosure with Local Preamplifiers
IMPORTANT:
The cable lengths for a given pair of transducers, including the preamplifier-to-electronics enclosure
cables, must be equal to within ±4 in. (10 cm) of each other.
For an XGF868i system configured with either a local or remote electronics enclosure and local preamplifiers, wire the
transducers and preamplifiers as follows:
1.
The local preamplifier must be installed in the junction box on the end of the transducer. Connect the BNC cable
with the mating connector from the preamplifier to the transducer BNC connector.
2.
Locate the two CH1 preamplifier-to-electronics enclosure cables and connect one to each of the two CH1
preamplifiers in the junction boxes, using the remaining preamplifier connections. Then, feed the free ends of the
cables through the selected conduit holes in the electronics enclosure.
3.
for the location (
J3
) of the connections for the Channel 1 transducers in the electronics
enclosure. Then, refer to
Figure 10
to wire the transducer to preamplifier and preamplifier to electronics enclosure
connections.
Note:
The
RED
cable leads are the SIG(+) leads and the
BLACK
cable leads are the RTN(-) leads.
Attention European Customers!
To meet CE Mark requirements, all cables must be installed as described in
Appendix A,
CE Mark Compliance
.
4.
For a 2-Channel XGF868i, repeat steps 1-3 to connect the CH2 transducers to terminal block
J4
in the electronics
enclosure. Note that it is not required that both channels of a 2-channel unit be connected.
5.
If wiring of the system has been completed, reinstall the rear cover and o-ring on the electronics enclosure and
tighten the set screw.
Note:
A channel must be activated in the
Channel Menu
before it can begin taking measurements. See
, for instructions.
WARNING!
Before connecting the transducers, take them to a safe area and discharge any
static buildup by shorting the center conductor of the transducer cables to the
metal shield on the cable connector.
Summary of Contents for DigitalFlow XGF868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 7: ...DigitalFlow XGF868i Startup Guide v Preface...
Page 8: ...Preface vi DigitalFlow XGF868i Startup Guide no content intended for this page...
Page 34: ...Chapter 1 Installation DigitalFlow XGF868i Startup Guide 24...
Page 46: ...Chapter 2 Initial Setup 36 DigitalFlow XGF868i Startup Guide...
Page 48: ...Chapter 2 Initial Setup DigitalFlow XGF868i Startup Guide 38...
Page 68: ...Appendix A CE Mark Compliance 58 DigitalFlow XGF868i Startup Guide...
Page 74: ...Appendix B Data Records 64 DigitalFlow XGF868i Startup Guide...
Page 80: ...Index 4 DigitalFlow XGF868i Startup Guide...
Page 82: ...Warranty 6 DigitalFlow XGF868i Startup Guide...
Page 84: ......
Page 86: ......
Page 88: ......
Page 89: ......