Chapter 5. Specifications
52
DigitalFlow™ XGF868i Startup Guide
5.1.7
Repeatability
5.1.7.1
High Flow Range: ±1.0 ft/s to ±394 ft/s (±0.3 to ±120 m/s)
One-Path:
±0.5 to 1.0%
Two-Path:
±0.35 to 0.75%
5.1.7.2
Low Flow Range: ±0.1 ft/s to ±1.0 ft/s (±0.03 to ±0.3 m/s)
One-Path:
±5 to 6%
Two-Path:
±3.5 to 4.0
5.1.8 Rangeability (Overall)
4000:1
5.2 Electronics
5.2.1
Flow Measurement
Patented
Correlation Transit-Time
mode
5.2.2 Enclosures
Standard:
Epoxy-coasted aluminum, Hazardous area ratings:
Explosion-proof:
Class I, Division 1, Groups B, C, D
Flameproof:
ISSeP 07ATEX015
II 2 G Ex d IIC T5 IP66
IECEx:
FM G 0011x
II 2 G Ex IIC T6 Gb IP66
Optional:
Stainless steel
5.2.3 Dimensions (l x d)
8.2 in. x 6.6 in. (208 mm x 168 mm)
5.2.4 Weight
Aluminum:
10 lb (4.5 kg)
Stainless steel:
30 lb (13.6 kg)
Summary of Contents for DigitalFlow XGF868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 7: ...DigitalFlow XGF868i Startup Guide v Preface...
Page 8: ...Preface vi DigitalFlow XGF868i Startup Guide no content intended for this page...
Page 34: ...Chapter 1 Installation DigitalFlow XGF868i Startup Guide 24...
Page 46: ...Chapter 2 Initial Setup 36 DigitalFlow XGF868i Startup Guide...
Page 48: ...Chapter 2 Initial Setup DigitalFlow XGF868i Startup Guide 38...
Page 68: ...Appendix A CE Mark Compliance 58 DigitalFlow XGF868i Startup Guide...
Page 74: ...Appendix B Data Records 64 DigitalFlow XGF868i Startup Guide...
Page 80: ...Index 4 DigitalFlow XGF868i Startup Guide...
Page 82: ...Warranty 6 DigitalFlow XGF868i Startup Guide...
Page 84: ......
Page 86: ......
Page 88: ......
Page 89: ......