020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
Function buttons
o
Select recipe
o
Test mode
o
Settings
o
Inflate
o
o
Reset
The operation of the machine is controlled by this main menu only. The active recipe is
executed by pushing the Start button. This starts the mixer (see the set acceleration &
deceleration times) until it reaches the set rpm and the mixing and alternation time start
running. As soon as the alternation time is reached the mixer changes the direction of
revolution. The mixer stops when the mixing time has elapsed. If so set, the data are printed
out on the printer.
If during mixing the Pause button is pushed the mixer decelerates in accordance with the set
deceleration time. The actual data remain valid so that the recipe can be continued with
the remaining actual values when the Start button is pushed again.
If the Stop button is pushed the mixer decelerates and the actual values are deleted.
In both cases the mixer always returns to the zero position. This is the position where
connecting the mixing biocontainer to or disconnecting the mixing biocontainer from the
system is effected most easily (with the paddle pointing towards the back of the mixing
tank).
If the mixing cycle is interrupted by an emergency stop the mixer will immediately stop and
the actual data will be deleted. The mixer will not stop in the zero position, but in this case
the mixing head can easily be manually turned to the zero position in order to disconnect the
mixing biocontainer.