020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
In most applications, the following sequence is used:
filling with liquids
connecting a powder handling bag to the mixing biocontainer (in case of powder
applications)
partially inflating the mixing biocontainer (in case the bag is less than 80% full)
connecting the mixing biocontainer to the system
introducing the powder by opening the powder handling bag
complete inflating of the mixing biocontainer
starting of the mixing
However, different applications or processes can have a different mixing biocontainer
handling sequence.
Make sure that the clamp on the drain tubing is closed and that the clamps on the filling
tubing are open. Connect the filling tube to the liquid supply and start filling.
It is important that the filling of the first liters of water happens gently to be able to position the
bottom square of the mixing biocontainer correctly, without too many wrinkles and stresses
on the film. After the first 10 - 20L the mixing biocontainer can be pulled gently towards the
four corners of the mixing tank.
Figure 30: Filling first 10-20L in mixing biocontainer