![Pall PadMixer 1000 Premium Instructions For Use Manual Download Page 26](http://html1.mh-extra.com/html/pall/padmixer-1000-premium/padmixer-1000-premium_instructions-for-use-manual_748037026.webp)
020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
5.3 Setting the mixing angle on the drive unit
The angle of the mixing stick can in theory be set between 5 and 25°. It is however not
recommended to use an angle above 12.5° in order to avoid damage to the mixing
biocontainer while mixing due to high stress on the film.
The angle can be changed by unlocking the 2 fixation pins at the back and by
pushing/pulling the mixing head in another angle while keeping the 2 other fixation pins in
the front unlocked.
Figure 32: Setting the mixing angle
The life time of a mixing biocontainer will decrease with increasing mixing angle as there are
more stresses on the mixing sleeve.
In general an angle of 10 or 12.5° will be more than sufficient to obtain an efficient mixture.
Therefore Pall strongly recommends to not use an angle above 12.5°.