020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
41
Now there is free access to the gearbox. Disconnect the power supply cable, remove the
RPM counter disc and the bracket with the 2 sensors. Disconnect the mixing head from the
motor axes and remove the motor from the support plate by means of the 4 screws.
The recommend oil type ISO VG 220.
6.7 Uploading a new PLC code
This chapter explains how a new or modified PLC code can be uploaded onto the systems
PLC.
When a new code is uploaded, some old information that is stored in the program will be lost.
Therefore note all Recipe settings and Factory settings so that these can be edited when the
new code is installed.
Load the new or modified PLC code on a CF-card (Compact Flash).
2 files + 1 folder should be loaded onto the CF-card, so not under another directory or folder.
The following type of files and folder should be saved on the CF-card :
Figure 47: Overview of files and folders to be saved on the CF card
Turn off the main power the mixing system.
Open the back of the touch screen box (4 screws). Open the CF-card slot and insert the CF-
card.
Figure 48: Inserting the CF card