Preparation for the calibration procedure by using a precise digital scale:
1) Connect the silicone tubing coming out from the pump #1 (test water sample outlet
coming from Connector 1 OUT) to the calibration stand connector (NO green tip).
2) Connect the second silicone tubing coming out from the pump #2 (reagent solution
outlet coming from Connector 3 OUT) to the calibration stand connector with the
green tip.
Initial Preparation Calibration steps:
1) Place the digital scale on the calibration stand.
2) Place a dedicated calibration cylinder (vial) on the digital scale.
3) Start
Channel #1
pump (Manual Control / Calibration Tab in the application). The
pump will start to pump the test water sample into the calibration vial. When you see
that the test water sample is coming out uniformly, stop the pump by clicking the
“
Manual button
” in the application.
4) Dispose of the fluid from the calibration cylinder (vial), and repeat, exactly the same
initial preparation calibration procedure for the reagent solution. Remember to run,
in this case
Channel #2
pump until the reagent solution is coming out from the
green tip uniformly.
Important!
Do NOT add back tested reagent solution from the calibration cylinder
(vial) to the reagent solution bottle/container!
Make sure that the calibration cylinder (vial) is empty, dry it with a clean paper towel and
place it again on digital scale.
Digital Scale