Page 47
Kore 7th Manual / kHLab Edition
DC AWC pump calibration procedure
1) Connect tubes to AWC pump set.
2) Immerse both pumps into water.
3) Activate the pump by pressing DC#1 and DC#2 to remove air from the pumps and
tubes.
4) Arrange an accurate measuring vessel and then insert the exit tube from DC 1 pump
into it.
5) Press DC#1 again, wait until proportioning is complete and then read amount of
measured water.
6) Amount of water shall be entered into Measured upper field [ml].
7) Evacuate the measuring vessel and place the tube from the other pump into it.
8) Press DC#2 again, wait until proportioning is complete, and read amount of
measured water.
9) Amount of measured water shall be entered into Measured lower field [ml].
10) Save the calibration by pressing
Save
settings