Mosaic3
57
Version 3.5 rev 8 May 2019
TECHNICAL SPECIFICATIONS
SECTION 8: TECHNICAL SPECIFICATIONS
8.1 ac/dc p
oWer
s
uppLy
r
equirements
Parameter Specification
Mains Input Voltage Range
100 – 240 V a.c. (typical)
Mains Input Frequency Range
43 – 67 Hz (typical)
Mains Input Steady State Current
0.1 A max. based on typical voltage range above
Low Voltage Output
12 V d.c. ± 5%
Low Voltage Output Ripple
200 mV max.
Low Voltage Steady State Current
Output
1 A max.
Low Voltage In-rush Current
Capability
Shall start up a load whose in-rush current from a 0.1 Ω source resistance is 21A min. peak
and a pulse width of 80 µs min. measured at half the peak
Low Voltage Output Connector
Tyco 3-1437719-3 4-pin Triad Female Cable Plug, which uses UL 94 V-0 insulators
Pins are shown looking into the connector from the insertion side:
General Safety Requirements
The power supply must be certified to an appropriate IEC standard, e.g. IEC 60950-1, and
meet the reinforced insulation from mains requirement of IEC 61010-1
Power Supply Overvoltage Category
CAT II. An overvoltage category of CAT II means that the power supply is designed to cope
with transient voltages above the rated supply that would be experienced by any product
connected to a mains socket in a building.
8.2 p
oWer
c
onsumption
Parameter Specification
Mosaic USB DMD Head + External
Power Supply
6W (typ. & max.)
Mosaic USB DMD Head Only
4.8W (typ. & max.)
8.3 e
LectricaL
s
pecifications
(i/o)
See
Section 3.1.7 Connector Signal Electrical Specification
.