Mosaic3
52
Version 3.5 rev 8 May 2019
OPERATION
5.7 c
hanging
the
d
ichroic
p
osition
(m
osaic
i
nfinity
)
The Mosaic Infinity Dichroic slider enables the selection between three positions. Move the Slider between the positions
as required- . These are defined in the adjacent label (Figure 22) to indicate which dichroic option is available at each
position. An example of the label is shown in Figure 23.
Figure 19: Dichroic Slider on Mosaic Infinity
Figure 20: Dichroic Slider Label
5.8 r
isk
m
itigation
5.8.1 i
nterLock
s
afety
s
ystem
t
esting
If a laser system is present then the operation of the Interlock safety system should be tested prior to use of the system.
When the system is operational, partially remove a Beam Splitter plug and ensure that the Emission LED on the front of
the diode lasers controller extinguishes.
If a fault is ever observed please report it immediately to your local Andor Support representative and stop all use of the
product.
5.8.2 m
echanicaL
h
ousing
Once installed, the Mosaic device and the microscope form the Protective Housing of the product. No components,
panels, connections or linkages should be loosened or removed to avoid exposure to hazardous radiation.
Move the Slider
between the available
positions as required